At a Glance

-Selling for parts/repairs

-Powers On
-Fails Initialization (the needle motor encoder might be faulty)


Condition Description
The HT3100 sample handler is in good cosmetic condition, but it is not in operational condition. The technicians tested this instrument for functionality and noted that it did not finish initializing. The homing needle exceeds the limits, suggesting the encoder needs replaced or repaired. The manufacturer was unable to troubleshoot the error, unless the technician purchased their training program.



Technical Evaluation

The needle exceeds limits when homing and initialization does not complete. The techniciains believe this error might be the result of a faulty motor encoder.