- Beckman Coulter Allegra X-30 centrifuge (B06314)
- Beckman Coulter SX4250 swing bucket rotor (392243)
- Four (4) Beckman Coulter SX4250 rotor buckets (392252)
- Four (4) Beckman Coulter 12x15mL bucket adapters (368331)
- Power cable
- 120-Day warranty
After its initial disinfection and inspection, our technicians tested the Beckman Coulter Allegra X-30 centrifuge with the following results:
- Ran to 4,200 rpm: PASSED
- No overheating condition occurred: PASSED
- Performed inspection, no problem found
- No excessive noise or imbalance
- Spun down properly
- Replaced power switch
- Replaced motor boot
- Replaced gas spring
- Repaired main board
Manufacturer
|
Beckman Coulter
|
Model
|
Allegra X-30
|
Cat. No.
|
B06314
|
Serial No.
|
ALT12F017
|
Mfg. date
|
2012
|
Maximum speed
|
16,000 rpm (rotor dependent)
|
Maximum g-Force
|
23511 x g
|
Acceleration
|
10 acceleration profiles
|
Deceleration
|
10 deceleration profiles
|
Time run
|
Up to 9 hrs, 59 mins: Continuous run (8)
|
Max capacity volume range
|
1600 mL
|
Refrigeration system
|
No
|
Drive type cooling
|
Brushless induction
|
Temperature range(s)
|
Ambient temperature range: +4° to +35°C
|
Friction reduction system
|
Asynchronous 3-phase direct drive
|
Installation category
|
II
|
Electrical requirements
|
120 V, 60 Hz, 6 A
|
Dimensions (W x D x H)
|
18.1 x 21.6 x 14 in
|
Weight
|
48 kg (105.8 lb)
|