- Beckman Coulter Optima MAX 130K ultracentrifuge
- Power cable
- 120-Day warranty
After its initial disinfection and inspection, our technicians tested the Beckman Coulter Optima MAX 130K ultracentrifuge with the following results:
- Ran to 130,000 rpm: PASSED
- Cooled down to 0°C during run: PASSED
- Vacuum pulled down to specification, below 10 microns: PASSED
- Performed inspection, no problem found: PASSED
- No excessive noise or imbalance
- Spun down properly
- Replaced tach sensor
- Vacuum seals cleaned and greased
- Diffusion, vacuum pump, and drive motor oil changed
- Main board upgraded
- Replaced bellows seal
Manufacturer
|
Beckman Coulter
|
Model
|
Optima MAX-130K
|
Catalog No.
|
364301
|
Serial No.
|
CTM04C01
|
Mfg. date
|
2004
|
Speed Range
|
5000 to 130000 rpm in 1000 RPM increments
|
Speed Control
|
Actual rotor speed, +/- 50 RPM of set speed
|
Speed Display
|
Actual rotor speed in 100 RPM increments above 5000 RPM and 10 RPM increments below 5000 RPM
|
Set Time
|
Up to 99 hours, 59 minutes
|
Time Display
|
Manual operation: indicates run time remaining Programmed operation: time remaining in step
|
Temperature Range
|
0 to 40°C in 0.1°C increments
|
Temperature Control (after equilibrium)
|
+/-2°C of set temperature
|
Temperature Display
|
Actual rotor temperature in 0.1°C increments
|
Ambient Temperature Range
|
15 to 35°C
|
Vacuum
|
below 10 microns (1.34 Pa)
|
Acceleration
|
10 acceleration profiles: 9 slow rates from 0 to 5000 RPM followed by full acceleration to set speed; or maximum acceleration
|
Deceleration
|
11 deceleration profiles: 10 slow rates from 5000 to 0 RPM, including coasting to a stop without brake; or full dynamic braking from set speed
|
Electrical Requirements
|
120V, 60Hz, 12A
|
Dimensions (W x D x H)
|
29 x 23 x 22 in
|