- Eppendorf 5430R Refrigerated Benchtop Centrifuge
- FA-45-30-11 Rotor with lid
After its initial disinfection and inspection, our technicians tested the 5430 with the following results:
- Runs to 14k rpms, refrigeration holds at -1°C, no excessive noise or imbalance. Spins down properly: PASSED
- Performed inspection. No problems found: PASSED
- Adjusted pins on door sensor connector
- Painted top lid and top back panel
Manufacturer
|
Eppendorf
|
Model
|
5430
|
Serial no.
|
|
Mfg. Year
|
2010
|
Max. Speed
|
17,500 rpm
|
Max. RCF
|
30,130 x g
|
Max. Rotor Capacity
|
48 x 1.5/2.0 mL, 6 x 50 mL, 2 x MTP
|
Temperature range
|
-11°C to 40°C
|
Power Requirements
|
120V, 50-60Hz, 12A, 1050W
|
Dimensions (W x D x H)
|
13" x 16.5" x 9.5"
|
Model No.
|
Type
|
Capacity
|
RPM / RCF
|
A-2-MTP
|
Microplate Swing-Bucket
|
x2 buckets
|
4,680 rpm / 2,204 x g
|
FA-45-24-11-HS
|
Fixed-Angle
|
24 x 1.5/2.0 mL tubes
|
17,500 rpm / 30,130 x g
|
F-45-30-11
|
Fixed-Angle
|
30 x 1.5/2.0 mL tubes
|
14,000 rpm / 20,817 x g
|
F-45-64-5-PCR
|
PCR Strip Tube
|
64 x 0.2 mL PCR tubes
|
11,800 rpm / 13,543 x g
|
F-35-6-30
|
Fixed-Angle
|
6 x 15/50 mL or 12 x Vacutainer tubes
|
7,830 rpm / 7,745 x g
|
FA-45-16-17
|
Fixed-Angle
|
16 x 5.0 mL Eppendorf tubes
|
14,200 rpm / 21,191 x g
|
FA-45-24-11-Kit
|
Fixed-Angle
|
24 x 1.5/2 mL spin columns
|
13,200 rpm / 19,090 x g
|
FA-45-30-11
|
Fixed-Angle
|
30 x 1.5/2 mL tubes
|
14,000 rpm / 20,817 x g
|
FA-45-48-11
|
Fixed-Angle
|
48 x 1.5/2 mL tubes
|
12,700 rpm / 18,210 x g
|
F-45-48-11
|
Fixed-Angle
|
48 x 1.5/2 mL tubes
|
12,700 rpm / 18,210 x g
|
S-24-11-AT
|
Swing-Bucket
|
24 x 1.5/3 mL tubes
|
12,700 rpm x 16,049 x g
|