- Eppendorf 5702 benchtop centrifuge
- Eppendorf A-4-38 swing bucket rotor
- Four (4) Eppendorf A-4-38 rotor buckets with inserts
- Power cable
- 120-Day warranty
After its initial disinfection and inspection, our technicians tested the Eppendorf 5702 benchtop centrifuge with the following results:
- Ran to 4,400 rpm: PASSED
- No overheating condition occurred: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Performed inspection, no problem found: PASSED
- Adjusted door latch microswitch
Manufacturer
|
Eppendorf
|
Model
|
5702
|
Serial No.
|
5702HJ945989
|
Mfg. date
|
April 2018
|
Max. RCF
|
3,000 × g
|
Max. RCF with fixed-angle rotor
|
1,900 × g
|
Max. RCF with swing-bucket rotor
|
3,000 × g
|
Display
|
LCD
|
Speed
|
100 – 4,400 rpm (100 rpm steps)
|
Max. Capacity
|
4 × 100 mL
|
Rotors available
|
6
|
Acceleration time
|
≤19 s
|
Deceleration time
|
≤19 s
|
Timer
|
0 – 99 min, with continuous run function
|
SOFT brake function
|
yes
|
Noise level
|
< 52 dB(A) with Rotor A-4-38
|
Volume range
|
400 mL
|
Electrical requirements
|
120V, 60Hz 2.3A
|
Max. power consumption
|
200 W
|
Dimensions (W × D × H)
|
12.6 × 15.6 × 9.6 in
|
Weight w/o accessories
|
19 kg / 41.9 lb
|
Cooling
|
non-refrigerated
|
Max. capacity:
|
Conical tubes
|
20 × 15 mL/ 4 × 50 mL
|
Blood collection tubes
|
36 × 13 mm/ 30 × 16 mm
|