Eppendorf epMotion 5070 Robotic Pipette Liquid Handler w/ 50uL Tool & Controller
Fully tested and working, this epMotion 5070 is guaranteed to meet performance specifications. Our technicians ensured that it successfully initialized without error and executed programs.
After its initial disinfection and inspection, our technicians tested this epMotion 5070 with the following results:
- Initializes without error: PASSED
- Connects to controller: PASSED
- Executes programs: PASSED
- Robotic arm is smooth: PASSED
- Sensors working: PASSED
- Aspiration works: PASSED
- Tip disposal works: PASSED
- Eppendorf epMotion 5070 Robotic Pipette Liquid Handler
- Controller
- Mouse
- 0-50uL Single-Channel Tool
- Open box of 50uL Pipette Tips
- Power and comms cables
Manufacturer
|
Eppendorf
|
Model
|
epMotion 5070
|
Serial No.
|
5070 01009
|
Operation
|
Electronic
|
Pipetting type
|
Air-cushion
|
SLAS/ANSI positions
|
4
|
Application
|
PCR/qPCR and normalization
|
Volume range
|
0.2 uL - 1 mL
|
Power supply
|
100-240V +/-10%, 50-60Hz +/-5%
|
Max. power consumption
|
150W
|
Dimensions (W x D x H)
|
26 x 19 x 25 in. (65 x 48 x 63 cm)
|
Weight w/o accessories
|
99.2 lbs (45 kg)
|
Optical sensor
|
Contact-free recognition of fill level, inserted tools, labware, tip types and quantities
|
Conductor — X, Y, Z positioning
|
Random measurement error: +/-0.1 mm Systematic measurement error: +/-0.3 mm
|
Dispensing tool
Note: All measurements taken in pipetting mode, free jet, without pre-wetting, with distilled water, at 20°C
|
Random measurement error (1 µL)
|
≤3 %
|
Random measurement error (1,000 µL)
|
≤0.4 %
|
Systematic measurement error (1 µL)
|
±5 %
|
Systematic measurement error (1,000 µL)1,2
|
±0.7 %
|
Systematic measurement error (50 µL)1,2
|
±1.2 %
|