- Leica CM1850 UV-3-1 cryostat microtome
- Blade holder assembly with anti-roll plate
- Two (2) specimen discs
- Power cable
- 120-day warranty
After its initial disinfection and inspection, our technicians tested the Leica CM1850 UV-3-1 cryostat microtome with the following results:
- Chamber temperature held at -32°C: PASSED
- Sample advance: PASSED
- Chamber light: PASSED
- Disinfection (UV) system: PASSED
- Winder heater: PASSED
- Handwheel brake: PASSED
- Fan alignment: PASSED
- Cleaned undercarriage
- Cleaned refrigeration condenser
- Cleaned chamber with microtome removed
- Cleaned and lubricated microtome
- Cleaned and lubricated knife assembly
Manufacturer
|
Leica
|
Model
|
CM1850 UV-3-1
|
Catalog No.
|
047140225
|
Serial No.
|
2057/10.2010
|
Mfg. date
|
10.2010
|
Cooling range
|
-35°C to 0°C
|
Section thickness range
|
1-60 µm From 1 to 10 µm in 1 µm intervals From 10 to 20 µm in 2 µm intervals From 20 to 60 µm in 5 µm intervals
|
Horizontal specimen feed
|
25mm / 0.98 in
|
Vertical stroke
|
59 mm / 2.32 in
|
Maximum specimen size
|
55 mm x 55 mm / 2.17 in x 2.17 in
|
Specimen orientation
|
8° (x-, y-, z-axis)
|
Disinfection time
|
30 or 180 min. cycle
|
Electrical requirements
|
120V, 60Hz, 1440VA
|
Dimensions (W x D x H)
|
29 x 28 x 47.5 in
|
Weight
|
~135 kg (296.62 lbs)
|