Manufacturer
|
Molecular Devices
|
Model
|
SpectraMax M2
|
Absorbance photometric performance
|
|
Wavelength selection
|
Monochromator, tunable in 1 nm increments
|
Wavelength range
|
200–1000 nm
|
Wavelength bandwidth
|
≤ 4.0 nm
|
Wavelength accuracy
|
±2.0 nm
|
Wavelength repeatability
|
±0.2 nm
|
Photometric range
|
0–4.0 OD
|
Photometric resolution
|
0.001 OD
|
Photometric accuracy (microplate)
|
< ±0.006 OD ±1.0%, 0–2 OD
|
Photometric accuracy (cuvette)
|
< ±0.005 OD ±1.0%, 0–2 OD
|
Photometric precision
|
< ±0.003 OD ±1.0%, 0–2 OD
|
Baseline flatness
|
< 0.001 OD
|
Stray light
|
< 0.05% @ 230 nm
|
Fluorescence photometric performance
|
|
Dual monochromators
|
1 nm increment selection
|
EX Wavelength range
|
250–850 nm
|
EM Wavelength range
|
360–850 nm
|
Wavelength bandwidth (EX, EM)
|
9 nm
|
Top-read optimized sensitivity
|
5 pM (fluorescein) in 96 wells
|
Top-read guaranteed* sensitivity
|
15 pM (fluorescein) in 96 wells
|
Time-Resolved Fluorescence (secondary mode)
|
|
EX Wavelength range
|
250-850 nm
|
EM Wavelength range
|
360–850 nm
|
Data collection
|
50–1450 µsec., 200 µsec. increments
|
Guaranteed sensitivity
|
7 pM Eu-chelate
|
Luminescence (secondary mode)
|
|
Wavelength range (M2)
|
360–850 nm
|
Guaranteed sensitivity
|
50 fM alkaline phosphatase
|
General photometric performance
|
|
Plate formats
|
6, 12, 24, 48, 96, 384 wells
|
Light source
|
Xenon flash lamp (1 joule/flash)
|
Detector
|
Photomultiplier tube (PMT)
|
Read time** 96 wells: 384 wells:
|
Abs 18 sec., FI 15 sec. Abs 49 sec., FI 45 sec.
|
Shaker time
|
0 to 999 seconds
|
Temperature control
|
Ambient +4°C to 45°C
|
Temperature uniformity
|
< 1°C at 37°C set point
|
Temperature accuracy
|
±1°C at 37°C set point
|
General specifications
|
|
Power Consumption
|
< 125 watts
|
Power
|
100–240 Vac, 3 A, 50/60 Hz
|
Dimensions
|
23x16x9”
|
Weight
|
35lbs
|