- Molecular Devices SpectraMax Plus 384 Absorbance Microplate Reader Spectrophotometer
- Dell Laptop with SoftMax Pro 6.1 Software Installed
- Laptop Charger
- Communication Cable
- Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Molecular Devices SpectraMax Plus 384 absorbance microplate reader spectrophotometer with the following results:
- Powers on, initializes, and connect to control software: PASSED
- Control panel and display: PASSED
- Ran spectrum scan of holmium oxide standard; results match published standards: PASSED
- Cuvette reading: PASSED
- Microplate reading: PASSED
- Plate loading/unloading: PASSED
Manufacturer
|
Molecular Devices
|
Model
|
SpectraMax Plus 384
|
Serial No.
|
MNR06804
|
Photometric performance
|
|
Wavelength range
|
190-1000 nm
|
Wavelength selection
|
Monochromator, tunable in 1.0 nm increments
|
Wavelength bandwidth
|
2 nm
|
Wavelength accuracy
|
±1.0 nm
|
Wavelength repeatability
|
±0.2 nm
|
Photometric range
|
0-4.0 OD
|
Photometric resolution
|
0.001 OD
|
Photometric accuracy (microplate)
|
<±0.006 OD ±1.0%, 0-2 OD
|
Photometric accuracy (cuvette)
|
<±0.005 OD ±1.0%, 0-2 OD
|
Photometric precision
|
<±0.003 OD ±1.0%, 0-2 OD
|
Baseline flatness
|
<0.001 OD
|
Stray light
|
<0.05% @ 230 nm
|
Light source
|
Xenon flash lamp
|
Microplate read time (normal mode)
|
96 wells in 9 seconds 384 wells in 29 seconds
|
Microplate read time (speed read mode)
|
96 wells in 5 seconds 384 wells in 16 seconds
|
Temperature regulation
|
|
Temperature range
|
Ambient +4°C to 45°C
|
Temperature uniformity (microplate)
|
±0.5°C at 37°C
|
General specifications
|
|
Robot compatible
|
Yes
|
Power rating
|
100-240V, 50-60Hz, 4A
|
Power consumption
|
<250 watts
|
Dimensions (W x D x H)
|
22.8 x 15 x 8.6 in (58 x 38 x 22 cm)
|
Weight
|
30 lb (13.6 kg)
|