After its initial disinfection and inspection, our technicians tested the NanoDrop 1000 UV/Vis spectrophotometer with the following results:
- Ran diagnostics check: PASSED
- Ran lamp intensity check: PASSED
- Spectrum scan; consistent, repeatable readings: PASSED
Manufacturer
|
NanoDrop Technologies
|
Model
|
ND-1000 UV/Vis
|
Serial No.
|
2875
|
Mfg. location
|
U.S.A.
|
Sample size
|
1 microliter
|
Path length
|
1 mm (with auto-ranging to 0.2 mm)
|
Light source
|
Xenon flash lamp
|
Detector type
|
2048-element linear silicon CCD array
|
Wavelength range
|
220-750 nm
|
Wavelength accuracy
|
1 nm
|
Wavelength resolution
|
3 nm (FWHM at Hg 546 nm)
|
Absorbance precision
|
0.003 absorbance (1mm path)
|
Absorbance accuracy
|
3% (at 0.74 absorbance at 350 nm)
|
Absorbance range
|
0.02 - 75 (10 mm equivalent absorbance)
|
Detection limit
|
2 ng/microliter (dsDNA)
|
Maximum concentration
|
3700 ng/microliter (dsDNA)
|
Measurement cycle time
|
10 seconds
|
Power input
|
12VDC, 6W max. consumption
|
Dimensions (W x D x H)
|
5.5 x 8 x 6 in
|