- Nuaire NU-S540-500 Class II Type A2 5ft biological safety cabinet
- Height adjustable stand
- Gas inlet valve
- 115 VAC power cable
- 120-day warranty
This biological safety cabinet does not include a HEPA filter. A HEPA filter will need to be purchased separately and can commonly be purchased by the company that will perform your IQ/OQ Certifications.
After its initial disinfection and inspection, our technicians tested the Nuaire NU-S540-500 Class II Type A2 5ft biological safety cabinet with the following results:
- Tested the blower sash location switch and alarms: PASSED
- Tested UV and white light: PASSED
- All control panel buttons work
- Interior outlets work
While this purchase does include the stand for the safety cabinet; the unit will not be shipped on the stand. This will ensure the unit will be able to pass through standard doorways at the point of delivery.
Manufacturer
|
Nuaire
|
Model
|
NU-S540-500
|
Serial No.
|
170235091515
|
Mfg. Year
|
2015
|
Air Cleanliness (ISO 14644)
|
ISO-Class-5
|
Airflow Monitor
|
Pressure
|
Bag In Bag Out
|
N/A
|
Cabinet Construction
|
All welded stainless steel 16/18 GA, Type 304 pressure tight design
|
Casters
|
N/A
|
Control System
|
AeroMax
|
Cord Pass Through
|
Standard Right, Optional Left
|
Data Output Features
|
RS-232
|
Decontamination Seal-able Window
|
N/A
|
Diffuser
|
Metal, non-flammable
|
Downflow fpm (m/s)
|
60 fpm (0.30 m/s)
|
Drain Valve Location
|
Front right
|
Electrical Configuration
|
115V 60Hz, 230V 50/60Hz
|
Front & Control Panel Material
|
Painted Steel
|
Fumigation: per NIH/NSF Procedures
|
TRUE
|
HEPA Filter Seal Type
|
HEPEX Zero Leak Neoprene Seal, Spring-loaded
|
Inflow fpm (m/s)
|
105 fpm (0.53 m/s)
|
IV Bar with 6 Hooks
|
Optional
|
Light Intensity fc (Lux)
|
90-120 (968-1291)
|
Motorized Base Stand
|
Optional as Separate Accessory
|
Motorized Window
|
N/A
|
Other Built-in Exhaust Option(s)
|
N/A
|
Other Built-in Supply Filter Option(s)
|
N/A
|
Outlet(s)
|
Standard, two backwall
|
Performance Standard
|
NSF-ANSI-49
|
Plenum Type
|
HEPEX
|
Power Cord Length Foot (Meter)
|
12 (3.7)
|
Pre-filter
|
N/A
|
Protection
|
Product, Personnel, Environmental
|
Reach into workzone Inches (mm)
|
6 (152)
|
Remote Alarm Terminals
|
Standard
|
Safety Certification
|
UL, UL-C
|
Style
|
Bench top, optional base stand
|
Ultraviolet (UV) Light
|
Optional, Backwall
|
Viewing Window Inches (mm)
|
21 (533) open to fully closed
|
Work Access Opening(s)
|
8, 10-inches (203, 254 mm)
|
Work Surface
|
One-Piece Dished Stainless Steel with Prop-Up Rod
|