- Perkin Elmer Victor X4 Multimode Microplate Plate Reader (2030-0040)
- Dell Laptop with Perkin Elmer 2030 Workstation Software Installed
- Laptop Charger
- USB 2.0 to ARCNET Adapter
- x2 Data Cables
- x2 Power Cables
- 90-Day Warranty
- Fluorescence intensity: top and bottom
- Luminescence: flash and glow
- Absorbance (UV and visible)
- Time-resolved Fluorescence
After its initial disinfection and inspection, our technicians tested the Perkin Elmer Victor X4 multimode plate reader with the following results:
- Reader initializes and connects to control software: PASSED
- Ran several absorbance scans on holmium oxide standard with appropriate results: PASSED
Manufacturer
|
Perkin Elmer
|
Model
|
Victor X4
|
Model No.
|
2030-0040
|
Serial No.
|
20300434
|
Luminescence Detection (96-well plate)
|
Lower limit of detection (LLD):
|
Total flux of 20,000 photons/s (standard PMT) 100,000 photons/s (red-sensitive PMT)
|
ALP
|
1 amol/well with AMPPD substrate
|
Flash luminescence
|
80 amol/well ATP
|
Glow luminescence
|
0.9 pg/well using britelite
|
Absorbance Detection (96-well plate)
|
Measuring range @ 405 nm
|
0-4 A
|
Accuracy @ 405 nm
|
<2% (or 0.01 A) within 0-2 A
|
Precision @ 405 nm
|
<0.5% (or 0.01 A) within 0-2 A
|
TR-Fluorescence Detection (200 uL, 96-well plate)
|
Europium
|
<6 amol/well, 30 fM
|
Linearity
|
>5 decades
|
Crosstalk
|
<0.01%
|
Terbium
|
<5 amol/well, 25 fM
|
Samarium
|
<50 amol/well, 250 fM
|
Dysprosium
|
<150 amol/well, 750 fM
|
Fluorescence Detection (200 uL, 96-well black plate)
|
Fluorescein
|
<2 fmol/well, 10 pM
|
Linearity
|
>5 decades
|
Crosstalk
|
<0.01%
|
Umbelliferone
|
<200 fmol/well, 1nM
|
Rhodamine
|
<100 fmol/well, 0.5nM
|
Physical Specifications
|
Power rating
|
110-120 V/220-240 V, 50/60Hz
|
Dimensions (W x D x H)
|
19.4 x 23.4 x 13.9 in
|
Weight
|
101.4 lb (46 kg)
|