- Thermo Scientific 3950 reach-in CO2 incubator
- Five (5) inner shelf doors
- Four (4) adjustable stainless steel shelves
- 115VAC power cable
- 120-day warranty
After its initial disinfection and inspection, our technicians tested the Thermo Scientific 3950 reach-in CO2 incubator with the following results:
- Calibrated temperature, reached and held steady at setpoint of 37°C: PASSED
- Calibrated CO2, opened source and equalized at setpoint: PASSED
Manufacturer
|
Thermo Fisher Scientific
|
Model
|
3950
|
Serial No.
|
318744-2458
|
Mfg. date
|
12/06/10
|
Electrical requirements
|
100/120V, 50/60Hz, 10A, 1PH
|
Dimensions (W x D x H)
|
Exterior: 38 x 32 x 80 in Interior: 30.5 x 26 x 60 in
|
Temperature
|
Control
|
+/- 0.1°C Microprocessor PID
|
Setpoint
|
Digital - Touch Pad, 0.1°C
|
Range
|
+5°C above ambient to 60°C
|
Uniformity
|
+/- 0.3°C @ +37°C
|
Tracking alarm
|
User programmable (low) indicator
|
Overtemp
|
Tracking, user programmable, action and indicator
|
Display
|
Digital, LED, 0.1°C increments
|
Safety type
|
Extreme temperature safety, action and indicator
|
Safety sensor
|
Thermostat, independent of temperature control system
|
Safety indicator
|
Message center, audible and visual alarms
|
Relative Humidity
|
Selectable ranges
|
Medium (>80%), High (>90%)
|
CO2
|
Control
|
+/- 0.1% microprocessor PID
|
Sensor
|
Thermal conductivity
|
Readability
|
0.1%
|
Range
|
0 to 20%
|
Inlet pressure
|
15 PSIG (+/- 5 PSIG)
|
Display
|
Digital LED, 0.1% increments
|
Shelves
|
Dimensions
|
30.62 W x 25.81" F-B (77.78 x 65.56 cm)
|
Construction
|
Solid stainless steel, 2B finish
|
Surface area
|
5.4 sq. ft (0.51 sq. m) per shelf
|
Max per chamber
|
145.8 sq. ft (13.55 sq. m)
|
Standard
|
5 (shipped with unit)
|
Maximum
|
27
|
Clearance
|
Adjustable on 2" (5 cm) centers
|
Fittings
|
Access port
|
2.4" (6.1 cm) ID, one port per side
|
CO2 inlet
|
1/4" barbed
|
Sample port
|
Front mount, 1/4" barbed
|
Water inlet
|
1/4" MPT for automatic fill
|
Overflow drain
|
3/8" MPT
|
Condensate drain
|
3/8" FPT
|