- Beckman Coulter Avanti J-20 XP Refrigerated Floor Centrifuge (368608)
- Beckman Coulter JS-5.3 Swing Bucket Rotor (368690)
- x4 Beckman Coulter JS-5.3 Rotor Buckets
- 4 Beckman Coulter JS-5.3 Bucket Adapters
- Power Cable
- 120-Day Warranty
After its initial disinfection and inspection, our technicians tested the Beckman Coulter Avanti J-20 XP refrigerated floor centrifuge with the following results:
- Ran to 5,300 rpm: PASSED
- Refrigeration held at 3°C: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Performed inspection; no problems found: PASSED
Manufacturer
|
Beckman Coulter
|
Model
|
Avanti J-20 XP
|
Cat. No.
|
368608
|
Serial No.
|
JXP05C13
|
Mfg. date
|
2005
|
Max. speed
|
26,000 rpm
|
Max. RCF
|
82,000 × g (JA 30.50 Ti Rotor)
|
Speed control
|
0-10,000 rpm: +/- 10 rpm 10,000-26,000 rpm: +/- 0.1% or equal
|
Refrigerant
|
R404A (Non CFC)
|
Maximum capacity
|
6 Liters
|
Single tube vol. range
|
1.5 mL - 1,000 mL
|
Set temperature range
|
-10° C to 40° C
|
Temperature control
|
± 2° C
|
User programs
|
30 programs, 2 steps (“I” version only)
|
Friction reduction system
|
Automatic
|
Max. heat dissipation to room
|
2.0 kw, 7,500 Btu/hr.
|
Noise level
|
57 dbA (0.91 M (3 ft.) in front of instrument at 20,000 rpm
|
Power rating
|
200/208/240V, 50/60 Hz, 24A
|
Dimensions (W x D x H)
|
28 x 33.5 x 45.5 in
|
Weight
|
639.4 lbs (290 kg)
|