After its initial disinfection and inspection, our technicians tested the Beckman Coulter Avanti J-25 refrigerated floor centrifuge with the following results:
 
- Ran to 20,000 rpm: PASSED
- Refrigeration held at -10°C: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Performed inspection; no issue found: PASSED
 
- Replaced Hall effect sensor.
 
| 
 Manufacturer 
 | 
 Beckman Coulter 
 | 
| 
 Model 
 | 
 Avanti J-25 
 | 
| 
 Cat. No. 
 | 
 363102 
 | 
| 
 Serial No.  
 | 
 JHY04F05 
 | 
| 
 Mfg. date 
 | 
 2004 
 | 
| 
 Speed range 
 | 
 100 to 25,000 rpm 
 | 
| 
 Setting range 
 | 
 in 100-rpm increments below 10,000 rpm in 500-rpm increments above 10,000 rpm or equivalent rcf (in 100 x g increments) 
 | 
| 
 Speed display 
 | 
 from 0 to 10,000 rpm, indicates actual rotor speed ±10 rpm from 10,000 to 25,000 rpm, indicates actual rotor speed ± 0.1%, or equivalent rcf 
 | 
| 
 Time setting range 
 | 
 to 180 min or continuous (hold) 
 | 
| 
 Time display 
 | 
 indicates run time remaining (timed run) or elapsed time (hold run) 
 | 
| 
 Temperature setting range 
 | 
 -20 to +40°C (in 1°C increments) 
 | 
| 
 Rotor temperature 
 | 
 ±2°C of set temperature (after equilibration) 
 | 
| 
 Ambient temperature range 
 | 
 16 to 38°C (60 to 100°F) 
 | 
| 
 Acceleration 
 | 
 Maximum or slow 
 | 
| 
 Deceleration 
 | 
 Maximum, slow, or off 
 | 
| 
 Power rating 
 | 
 200/208/240V, 50/60Hz, 24A 
 | 
| 
 Dimensions (W x D x H) 
 | 
 28 x 34 x 45.5 in 
 | 
| 
 Weight 
 | 
 583 lb (264.5 kg) 
 |