- Beckman Coulter Avanti J-E High-Speed Refrigerated Floor Centrifuge (369001)
- Beckman Coulter J-Lite Series JLA-10.500 Fixed-Angle Rotor and Lid
- x6 Beckman Coulter JLA-10.500 Biosafety Liquid Canisters
- Attached Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Beckman Coulter Avanti J-E high-speed refrigerated floor centrifuge with the following results:
- Ran to 10,000 rpm: PASSED
- Refrigeration held at 1°C: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Performed inspection; no issue found: PASSED
|
Manufacturer
|
Beckman Coulter
|
|
Model
|
Avanti J-E
|
|
Catalog No.
|
369001
|
|
Serial No.
|
JSE11G16
|
|
Mfg. date
|
July 15, 2011
|
|
Maximum speed
|
21,000 rpm
|
|
Maximum RCF
|
53,300 x g
|
|
BioSafe
|
Yes
|
|
Maximum capacity volume range
|
1 - 4 L
|
|
Timed run
|
1 min to 99:59 hrs, Hold
|
|
Set temperature range
|
-10°C to 40°C in 1° increments
|
|
Temperature control
|
±2°C of set temperature after equilibration
|
|
Accel/Decel profiles
|
2/3
|
|
Zonal capability
|
Yes
|
|
Refrigeration system
|
Non-CFC, non-ozone depleting refrigerant
|
|
Drive type cooling
|
Air-cooled
|
|
Power rating
|
208/240V, 60Hz, 24A
|
|
Dimensions (W x D x H)
|
25 x 33 x 36 in (63.5 x 80 x 91.4 cm)
|
|
Weight
|
589.5 lb (267.4 kg)
|
|
Manufacturer
|
Beckman Coulter
|
|
Type
|
Fixed-Angle Rotor
|
|
Model
|
JLA-10.500
|
|
Serial No.
|
11U1416
|
|
Mfg. date
|
06/11
|
|
Maximum speed
|
10,000 rpm
|
|
Max g-Force
|
18,600 x g
|
|
k-Factor
|
2,850
|
|
Nominal capacity
|
3000 mL
|
|
rMAX
|
166
|
|
rMIN
|
54
|
|
rAV
|
110 mm
|
|
BioSafe
|
Biosafety liquid canisters
|
|
Materials
|
Anodized aluminum
|