- Beckman Optima MAX 130K Benchtop Ultracentrifuge (364301)
- Beckman Coulter Optima TLX Instruction Manual
- Beckman Ultracentrifuge Rotor Guide
- Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Beckman Optima MAX 130K benchtop ultracentrifuge with the following results:
- Ran to 120,000 rpm: PASSED
- Refrigeration held at 4°C: PASSED
- Vacuum pulled down to acceptable levels: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Performed inspection; no problem found: PASSED
- Replaced bellows seal.
- Changed vacuum and diffusion oil.
- Topped off drive oil.
- Cleaned and greased seals.
Manufacturer
|
Beckman Coulter
|
Model
|
Optima MAX (130K)
|
Cat. No.
|
364301
|
Serial No.
|
CTM99B03
|
Mfg. date
|
1999
|
Speed range
|
5000 to 130000 rpm in 1000 rpm increments
|
Speed control
|
Actual rotor speed, +/- 50 rpm of set speed
|
Speed display
|
Actual rotor speed in 100 rpm increments above 5000 rpm and 10 rpm increments below 5000 rpm
|
Set time
|
Up to 99 hours, 59 minutes
|
Time display
|
Manual operation: indicates run time remaining Programmed operation: time remaining in step
|
Temperature range
|
0 to 40°C in 0.1°C increments
|
Temperature control (after equilibrium)
|
+/-2°C of set temperature
|
Temperature display
|
Actual rotor temperature in 0.1°C increments
|
Ambient temperature range
|
15 to 35°C
|
Vacuum
|
below 10 microns (1.34 Pa)
|
Acceleration
|
10 acceleration profiles: 9 slow rates from 0 to 5000 RPM followed by full acceleration to set speed; or maximum acceleration
|
Deceleration
|
11 deceleration profiles: 10 slow rates from 5000 to 0 RPM, including coasting to a stop without brake; or full dynamic braking from set speed
|
Electrical requirements
|
120V, 60Hz, 12A
|
Dimensions (W x D x H)
|
29 x 23 x 22 in
|
Weight
|
217 lbs (99 kg)
|