- SP GeneVac EZ-2 4.0 Elite Centrifugal Evaporator EZ4T-12060-HN0 (100023738)
- Edwards nXDS6i 100/240V Dry Scroll Pump (A73501983)
- x2 GeneVac Rotor Buckets
- x2 GeneVac Vial Blocks (10-5013)
- Waste Jug
- Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the SP GeneVac EZ-2 4.0 Elite centrifugal evaporator with the following results:
- System self test: PASSED
- Pump vacuum: PASSED
- Condenser temperature: PASSED
- Rotor speed: PASSED
- Lamp life: PASSED
- Sample temperature control: PASSED
- Ambient temperature: PASSED
- Chamber heating: PASSED
- Drain valve pressure drop: PASSED
- Vent valve pressure drop: PASSED
- Vacuum leaks: PASSED
Note: The EZ-2 4.0 Elite is recommended for solvents with boiling points up to 202°C and suitable for evaporating high boiling point solvents and lyophilizing aqueous based samples.
Manufacturer
|
SP GeneVac
|
Model
|
EZ4T-12060-HN0
|
Part No.
|
100023738
|
Serial No.
|
EZ5246
|
Mfg. date
|
March 2023
|
Max. rotor speed
|
1920 rpm
|
Nominal sample load g force
|
500 g
|
Temperature control range
|
30°C to 80°C
|
Max. load per swing
|
1.5 kg
|
Max. operational imbalance
|
80 g
|
Ultimate system vacuum
|
<0.4 mbar
|
Condenser type
|
Single-stage vapor compressor
|
Refrigerant gas
|
R1270
|
Refrigerant charge
|
28 g
|
Refrigerant GWP
|
2
|
Refrigerant CO2e
|
<0.1 tons
|
Ultimate low temperature
|
-50°C
|
Max. pressure (PS)
|
30 bar
|
Max. solvent capacity
|
750 ml
|
Refrigeration ACC range
|
110°C
|
Noise (@ 1 meter)
|
65 dB(A)
|
Exhaust hose
|
6 mm ID / 8 mm OD
|
Power rating
|
120V, 60Hz
|
Power consumption
|
1220VA max.
|
Dimensions (W x D x H)
|
24.1 x 25.5 x 22 in (61.3 x 64.8 x 56 cm)
|
Weight approximate
|
176 lb (80 kg)
|
Manufacturer
|
Edwards
|
Model No.
|
nXDS6i 100/240V
|
Part No.
|
A73501983
|
Serial No.
|
230132586
|
Mfg. date
|
2023
|
Weight
|