- Avantik QS12 Cryostat Microtome (956660)
- Blade Holder Assembly with Anti-Roll Plate
- Heat Extractor
- x4 Specimen Discs
- Instruction Manual
- Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Avantik QS12 cryostat microtome with the following results:
- Chamber temperature holds -35°C (manufacturer’s spec): PASSED
- Checked sample advance: PASSED
- Checked chamber light: PASSED
- Checked window heater: PASSED
- Verify handwheel brake: PASSED
- Checked fan alignment and service: PASSED
- Cleaned undercarriage.
- Cleaned refrigeration condenser.
- Cleaned chamber with microtome removed.
- Cleaned and lubricated microtome and knife assembly.
Manufacturer
|
Avantik Biogroup LLC
|
Model
|
QS12
|
Ref. No.
|
956660
|
Serial No.
|
S1802080
|
Cryobar cooling (peltier element)
|
Max. -55°C (-67°F)
|
Chamber temperature
|
Max. -35°C -2K/+1K (-31°F -2K/+1K) at +20°C room temp. (decrease 0,7°C/1°C room temp. increase)
|
Defrosting options
|
Programmed/Immediate
|
Section thickness range
|
1-500 µm
|
Fine section thickness range
|
1-100 µm
|
Fine section thickness resolution
|
1-10 µm in 1 µm steps 10-50 µm in 2 µm steps 20-50 µm in 5 µm steps 50-100 µm in 10 µm steps
|
Trimming thickness range
|
5-500 µm
|
Trimming thickness resolution
|
5-30 µm in 5 µm steps 30-100 µm in 10 µm steps 100-200 µm in 20 µm steps 200-500 µm in 50 µm steps
|
Specimen retraction (return travel)
|
20 µm
|
Vertical specimen stroke
|
64 mm
|
Horizontal specimen movement
|
28 mm
|
Specimen approach
|
Automated via arrow buttons
|
Chuck size
|
30, 40 mm (additional special sizes are available)
|
Specimen orientation
|
x - and y - axes universal 8° with zero positioning; z-axis up to 360°
|
Coarse feed
|
Motorized
|
Cutting drive
|
Manual via handwheel
|
Handwheel lock
|
Mechanical
|
Clearance angle adjustment
|
Disposable blade carrier 8-16°
|
Temperature (operating limits)
|
+5°C to +35°C (+41°F to +95°F) (at a max. Relative humidity of 60%)
|
Temperature (recommended operation)
|
+15°C to +30°C (+59°F to +86°F) Note: Performance may deteriorate when operated outside of this range.
|
Temperature (transport and storage)
|
-20°C to +50°C (-4°F to 122°F)
|
Relative humidity
|
Max. 60% RH up to 35°C
|
Altitude
|
Up to 2000 m (6,500 ft)
|
Pollution degree
|
2
|
Over voltage category
|
II
|
Floor loading requirements
|
230 kg/m2 (507 lb/ft2)
|
Protection class
|
I
|
Sound emission
|
45 dB (A) - measured with 1m distance to the instrument.
|
Refrigerant
|
R404a, filling amount: 100...120V 172g
|
Power rating
|
100-120V~, 60Hz, 7A
|
BTU
|
1020 BTU/hour
|
Dimensions (W x D x H) w/o handwheel
|
25.19 x 29.9 x 45.3 in (64 x 76 x 115 cm)
|
Weight
|
315.2 lb (143 kg)
|