- Leica CM1950 Cryostat Microtome (14047742464) with Object Cooling and UV Disinfection
- Blade Holder Assembly with Anti-Roll Plate
- x2 Specimen Discs
- Heat Extractor
- Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Leica CM1950 cryostat microtome with the following results:
- Chamber temperature reaches -32°C: PASSED
- Object cooling temperature reaches -47°C: PASSED
- Check sample advance: PASSED
- Check chamber light: PASSED
- Check UV disinfection system: PASSED
- Check window heater: PASSED
- Verify handwheel brake: PASSED
- Cleaned undercarriage
- Cleaned refrigeration condenser
- Cleaned chamber, microtome, and knife assembly
- Lubricated microtome and knife assembly
Manufacturer
|
Leica
|
Model
|
CM1950
|
Reference No.
|
14047742464
|
Serial No.
|
00000006343
|
Mfg. date
|
2017-03
|
Power rating
|
120V~, 60Hz
|
Power consumption
|
1900VA
|
Dimensions (W x D x H)
|
32.8 x 33.5 x 47.8 in (83.5 x 85 x 121.5 cm)
|
Weight without accessories
|
320-425 lb (145-193 kg) depending on configuration
|
Microtome
|
Type
|
Rotary microtome, encapsulated
|
Section thickness range
|
1 to 100 um
|
Trimming range
|
Clinic: 10 - 40 um Research: 1 - 600 um
|
Total specimen feed
|
25 mm + 1 mm
|
Vertical stroke
|
59 mm +/- 0.5 mm
|
Specimen retraction
|
20um (can be deactivated)
|
Maximum specimen size
|
50 x 80 mm
|
Cutting speed
|
Slow: 0-50 strokes/min Fast: 0-85 strokes/min Vmax: 85-90 strokes/min
|
Specimen orientation
|
8° (x, y-axis)
|
Electric coarse feed
|
Slow: 300 um/s Fast: 900 um/s
|
Object Head Cooling
|
Temperature range
|
-10°C to -50°C
|
Defrosting
|
Electrical heater
|
Automatic defrosting
|
No
|
Manual defrosting
|
Yes
|
Defrost duration
|
15 minutes
|
Defrost temperature
|
45°C
|
Refrigeration System (60Hz)
|
Temperature range
|
0°C to -35°C +/-5 K, adjustable in 1 K increments at ambient temperature of 20°C
|
Cooling time to -25°C
|
approx. 5 hours
|
Cooling time to -35°C
|
approx. 8 hours
|
Refrigeration capacity
|
690 W
|
Cut-out pressure
|
25 bar
|
Safety factor
|
3
|
General Information
|
Operating temperature range
|
18°C to 35°C
|
Temperature range during storage
|
+5°C to +55°C
|
Relative humidity
|
max. 60% non-condensing
|
Storage humidity
|
<60%
|