- Beckman Coulter Avanti J-26 XP Refrigerated Floor Centrifuge (393124)
- Attached Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Beckman Coulter Avanti J-26 XP refrigerated floor centrifuge with the following results:
- Ran to 20,000 rpm: PASSED
- Refrigeration held at -6°C: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Performed inspection; no issue found: PASSED
Manufacturer
|
Beckman Coulter, Inc.
|
Model
|
Avanti J-26 XP
|
Cat. No.
|
393124
|
Serial No.
|
JXS07H49
|
Mfg. date
|
2007
|
Set speed range
|
100 to 26,000 rpm (in 100-rpm increments below 10,000 rpm, 500-rpm increments above 10,000 rpm) or equivalent RCF (in 100 x g increments)
|
Elutriation speed setting range
|
0 to 5,000 rpm (in 10-rpm or 10 x g increments)
|
Speed display
|
from 0 to 10,000 rpm, display indicates actual rotor speed +/- 10 rpm from 10,000 to 26,000 rpm, indicates actual rotor speed +/- 0.1%, or equivalent RCF
|
Time setting range
|
to 180 min or continuous
|
Time display
|
indicates run time remaining (timed run) or elapsed time (hold run)
|
Set temperature range
|
-10° to +40°C (in 1°C increments)
|
Temperature accuracy
|
rotor temperature controlled to within +/-2°C of set temperature (after equilibration)
|
Ambient temperature range
|
16 to 36°C (60 to 100°F)
|
Cooling fluid
|
Refrigerant 404A (HFC)
|
Acceleration
|
maximum or slow
|
Deceleration
|
maximum, slow, or off
|
Power rating
|
200/208/240V, 50/60Hz, 24A
|
Dimensions (W x D x H)
|
28 x 34 x 45.5 in
|
Weight
|
640 lb (290 kg)
|