- SP Scientific VirTis 50L Ultra EL-85°C Pilot Clean Room Configuration Freeze Dryer Lyophilizer (100012485)
- UPS Battery Backup
- Computer (with Monitor, Keyboard, Mouse) and LyoS 2.0 Control Software
- Communication Cable
- Attached Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the SP Scientific VirTis 50L Ultra EL-85°C Pilot-Production clean room freeze dryer with the following results:
- Condenser temperature achieved -84.1°C: PASSED
- Shelf temperature achieved -72.4°C: PASSED
- Vacuum achieved 16mT: PASSED
Manufacturer
|
SP Scientific
|
Model
|
50L Ultra EL-85°
|
Part No.
|
100012485
|
Serial No.
|
322291
|
Lowest shelf temperature (50Hz / 60Hz)
|
≤ -67°C / -70°C
|
Shelf temperature control range
|
-55 to 65°C
|
Shelf pull-down from 20°C to -40°C
|
≤ 40 minutes
|
Lowest condenser temperature (50Hz / 60 Hz)
|
≤ -82°C / -85°C
|
Maximum condenser capacity
|
≥ 50 L
|
Condenser surface area
|
10 ft2 (.93 m2)
|
Condenser pull-down from 20°C to -45°C
|
≤ 35 minutes
|
Maximum ice condensing capacity in 24 hours
|
≥ 20 L
|
Maximum deposition rate
|
≥ 0.83 L/hour
|
Number of compressors
|
2
|
Compressor horsepower
|
3.5 hp, 1.5 hp
|
Vacuum time to 100 mTorr
|
≤ 45 minutes
|
Vacuum rate of rise
|
≤ 30 mT/hour
|
Vacuum control +/- 5 mTorr
|
50mTorr to 500mTorr
|
Volume-based leak rate
|
≤ .0042 mbar L/sec
|
Lowest system vacuum
|
≤ 15 mT
|
Temperature uniformity
|
± 1.0°C
|
Construction
|
316L stainless steel shelves, product chamber and condenser chamber
|
Stoppering
|
Bottom-up hydraulic
|
Defrost type
|
Hot gas
|
Refrigerant type
|
CFC-free
|
Vapor port
|
8 in (20.3 cm)
|
Power rating
|
208-230VAC, 60Hz, 50A, 1Ph 220VAC, 50Hz
|
Recommended outlet
|
NEMA 6-50R
|
Overall dimensions (W x D x H)
|
25 x 78 x 75.75 in (64 x 198 x 192 cm)
|
Minimum clearance
|
10 in (25.4 cm)
|
Weight
|
2,000 lb (909 kg)
|
Note: Performance specifications are based on SP test data from units operating at an ambient room temperature of approximately 20°C. SP recommends an optimum operation range of 15-25°C (59-77°F).
Note: SP recommends a 24-inch (61 cm) clearance around all sides of the unit for serviceability. If machines are placed side by side, increase the minimum clearance to 48 inches (121.9 cm)
|
Air-Cooled
|
Water-Cooled
|
Compressed air (for units with isolation valve)
|
80 psig (5.5 bar)
|
80 psig (5.5 bar)
|
Ambient room temperature
|
15-25°C (59-77°F)
|
15-25°C (59-77°F)
|
Approx. peak room heat generated
|
22,900 BTU/h (6.7 kW)
|
5,400 BTU/h (1.6 kW)
|
Cooling water usage
|
N/A
|
2-5 gpm (8-19 Lpm)
|
Cooling water load
|
N/A
|
17,500 BTU/h (5.1 kW)
|
|
Gas 1
|
Gas 2
|
F gas
|
R407C
|
R508B
|
Charge (Kg)
|
1.245
|
0.970 (AC); 0.452 (WC)
|
GWP
|
1774
|
13396
|
EPA SNAP
|
IP/VLTR
|
IPR/VLTR
|
Safety Class
|
A1
|
A1
|
Total CO2e (t)
|
15.198 (AC); 8.264 (WC)
|
15.198 (AC); 8.264 (WC)
|