- Thermo Scientific CryoMed 7454 1.7 cu. ft. 120V CRF Controlled-Rate Freezer
- Serial Cable
- Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Thermo Scientific CryoMed 7454 controlled-rate freezer with the following results:
- Powers on and initializes: PASSED
- Control panel and buttons: PASSED
- Temperature accuracy: PASSED
- Solenoids engage: PASSED
- Fan spins: PASSED
Notice: This freezer has some cosmetic blemishes. Please review the available photos to see the physical condition before purchasing.
|
Manufacturer
|
Thermo Fisher Scientific
|
|
Type
|
Controlled-rate freezer
|
|
Product line
|
CryoMed
|
|
Model
|
7454
|
|
Serial No.
|
300303169
|
|
Mfg. date
|
20190801
|
|
Capacity
|
1.7 cu. ft. (48L)
|
|
Chamber temperature range
|
+50°C to -180°C
|
|
Chamber freezing rate
|
Minimum 0.1°C/min Maximum 99.9°C/min over operating temperature range Note: Rates vary based on load, LN2 pressure, unit size, and voltage.
|
|
Chamber warming rate
|
Minimum of 0.1°C/min (no load, T-couple in free air, center of chamber) Maximum of 12.0°C/min (same conditions as above) Average of 9.0°C/min (same conditions as above, range -40°C to -12°C) Note: Rates vary based on load, LN2 pressure, unit size, and voltage.
|
|
Automatic restart
|
The microprocessor retains all programming in non-volatile memory. In the event of a power outage, the unit restarts automatically once power is restored.
|
|
Construction
|
Interior ..... Stainless steel with removable ductwork for ease of cleaning Exterior ..... Stainless steel with powder coated finish Insulation ... Foamed in place polyurethane door and cabinet
|
|
Air flow
|
Horizontal from right to left, as viewed from the front
|
|
Sound level
|
Unit in operation does not exceed 65 dBA
|
|
Operational features
|
Door interlock switch turns off LN2, heater and fan motor, except during the warm-up cycle. During the warm-up cycle, the door can be opened with heater and fan running (up to 0°C) to accelerate warming. LN2 pressure relief valve located adjacent to LN2 connection. Chamber pressure vent port located on left side-wall. Heater overtemperature protection: At a temperature level of 93°C, an independent manual reset device interrupts power to the heating element.
|
|
External connections
|
Power entry PC interface port LN2 connection ½” flare
|
|
Plug type
|
NEMA 5-15P
|
|
Power rating
|
120V, 60Hz, 12A, 1Ph
|
|
Interior dimensions (W x D x H)
|
19 x 13 x 12 in
|
|
Exterior dimensions (W x D x H)
|
41.2 x 20.5 x 21 in
|
|
Weight
|
180 lb (81.6 kg)
|