- Applied Biosystems 9902 Veriti 96-Well PCR Thermal Cycler (4375786)
- 96-Well Block
- Power Cable
- 90-Day Warranty
 
After its initial disinfection and inspection, our technicians tested the Applied Biosystems 9902 Veriti 96-well thermal cycler with the following results: 
 
- Power and initialization: PASSED
- Ran between 95°C and 10°C for several hours: PASSED
 
| Manufacturer | Applied Biosystems | 
| Model | Veriti 96-Well Thermal Cycler | 
| Model No. | 9902 | 
| Part No. | 4375786 | 
| Serial No. | 299022165 | 
| Mfg. date | March 2008 | 
| Format | 0.2 ml tubes, 96-well plate | 
| Reaction volume range | 10-100 µL | 
| Block format | Non-interchangeable | 
| Block configurations | 96-well, 0.2 mL, 6-zone VeriFlex | 
| Temperature range | 4.0 to 99.9°C | 
| Thermal accuracy | ±0.25°C (35°C to 99.9°C) | 
| Thermal uniformity | <0.5°C (20 sec. after reaching 95°C) | 
| Max. ramp rate | 3.9°C/s (Block), 3.35°C/s (Sample) | 
| Reaction speed | Fast, Standard | 
| Capacity | 96 x 0.2 ml tubes, 1 x 96-well plate | 
| Display type | 6.5 in. VGA 32 k Color touchscreen | 
| User interface | USB | 
| Power rating | 100-240V~, 50/60Hz, 10A max. | 
| Dimensions (W x D x H) | 9.3 x 19.1 x 9.6 in (23.7 x 48.5 x 24.5 cm) | 
| Weight | 25 lb (11.4 kg) |