- Thermo Scientific TSQ Endura Triple-Stage Quadrupole Mass Spectrometer 230V
- Thermo Scientific Vanquish Column Compartment/Oven H (VH-C10-A)
- Thermo Scientific Vanquish Split Sampler HT (VH-A10A) (6850.1915)
- Thermo Scientific Vanquish Binary Pump H Quaternary (VH-P10-A)
- Windows Workstation with Monitor, Keyboard, and Mouse
- Divert/Inject Valve Module with Cables
- Chemyx Fusion 101 Syringe Pump
- Peak Scientific NM32LA Nitrogen Generator 230V
- Sogevac SV65 BIFC Pump
- Ethernet Cables
- Power Cables
- 90-Day Warranty
 
Notice: The above list is an overview and does not include everything that comes with this system. Continue reading to see a complete breakdown of the included items and a section on testing performed to verify functionality.
 
- Thermo Scientific Xcalibur
- Thermo Scientific TraceFinder
- Thermo Scientific Tune/Instrument Control
- Chromeleon (installed but license is required for operation)
 
- TSQ Endura mainframe (serial: TQH E1 0464) — triple-quadrupole MS
- API source options: H-ESI / HESI-II, and APCI (includes ion transfer tube + ion sweep cone)
- Divert/Inject valve module with cables
- Syringe pump (for infusion/tuning) with power supply (Chemyx fusion 101)
- MS power cord(s) (230 Vac), region-specific
- Ethernet cable(s), Cat5e shielded, ≤3 m (for PC/control)
- Collision gas line from MS to argon regulator + fitting
- Nitrogen gas line from source to source interface + fitting
 
- Vanquish Split Sampler module (cat. no. 6850.1915)
- Pump module: Quaternary 
- Column compartment/oven
- System interconnect cables (CAN/ethernet/control for Vanquish stack).
- LC power cords
- Bottle tray with caps & frits
- Autosampler racks/trays: 54 x 12 mm OD vials ( 22641.5 mL), 96 x 1.6, 7 and 8 mm OD vials ( 22641.2 mL), 24 x 15 mm OD vials ( 22644 mL), Deep well plates (96 and 384)
- Start/ready contact-closure cable (LC↔MS), if not using software handshake
 
- UHP & tubing to MS.
- Nitrogen (source/aux gas), HP, with PFA line to MS
- Lab exhaust for the forepump outlet and solvent waste trap vents
- Electrical outlets (typical):
- TSQ Endura: 230 Vac, ~5 A (1 outlet)
- Forepump: 230 Vac, ~8 A (1–2 outlets, 1 pump for Endura)
- PC/monitor/switch: 100–240 Vac (separate outlets)
- LC stack: per module (120/230 Vac). Total outlets commonly ≥10 for a single-forepump build
 
- Peek or stainless steel capillaries for LC→MS transfer
 
- Syringes (Hamilton) for direct infusion
- Waste containers/tubing with compatible caps
 
- Spare ion transfer tube & sweep cone (Endura)
- Spare forepump oil + exhaust mist filter
- Replacement autosampler needles/seals and sample loops (Vanquish Split Sampler)
- Pump seals/check valves (LC)
- Fuses (MS/LC)
 
- Autosampler confirmed functional through manual injection: PASSED
- LC system confirmed functional; flow path and divert valve operating correctly: PASSED
- Mass Spectrometer confirmed functional; capable of detecting analytes: PASSED
- Method Development substantially completed; ready for use in automated runs: PASSED
 
Notice: Continue reading for more details of our testing performed to verify functionality.
 
- A manual injection was performed to test the autosampler’s functionality.
- The needle was observed dipping correctly into the sample solution.
- Injection confirmed that the autosampler syringe system was operational and capable of aspirating/dispensing liquid.
 
- The LC pump and flow system were engaged and confirmed to be functional.
- Flow through the column and divert valve was verified, showing that the sample path to the mass spectrometer was intact.
- No leaks or pressure irregularities were observed, confirming the LC system’s stability.
 
- The TSQ Endura mass spectrometer was engaged and tested in operating mode.
- Signal response confirmed that the MS was detecting and recording ion events appropriately.
- This validated that the instrument was capable of receiving and processing analyte input from the LC system.
 
| Manufacturer | Thermo Fisher Scientific | 
| Model | TSQ Endura | 
| Serial No. | TQH-E1-0464 | 
| Mass range | m/z 10-3400 | 
| Resolution | Q1 and Q3 adjustable to 0.4 Da peak width (FWHM) across the entire mass range | 
| Scan rate | 15,000 amu/second at a resolution of 2 FWHM500 SRM/second (for any resolution from 0.4 through 2.0 FWHM)
 25msec polarity switching
 | 
| Mass stability | Mass assignment will be within ±0.05 Da over a 24 hour period. The laboratory room temperature must be maintained between 18–27 °C (65–81 °F). The room temperature may not change by more than 5 °C (9 °F) during this period. | 
| Positive Electrospray (HESI) | A 2 µL injection of a 500 fg/µL reserpine solution will produce a minimum signal-to-noise ratio of 80,000:1 for the transition of the protonated molecule at m/z 609.3 to the fragment ion at m/z 195.1 when operated in selected-reaction monitoring (SRM) mode with Q1 and Q3 resolution set to 0.4 and 0.7 Da FWHM respectively. | 
| Atmospheric Pressure Chemical Ionization (APCI) | A 2 μL loop injection of a 500 fg/μL reserpine solution will produce a minimum signal-to-noise ratio of 25,000:1 for the transition of the protonated molecular ion at m/z 609.3 to the fragment ion at m/z 195.1 when operated in selected reaction monitoring (SRM) mode with Q1 and Q3 resolution both set to 0.7 Da FWHM. | 
| Negative Electrospray (nESI) | A 2 μL loop injection of a 500 fg/uL chloramphenicol solution will produce a minimum signal-to-noise ratio of 80,000:1 for the transition of the deprotonated molecular ion at m/z 321.0 to the fragment ion at m/z 152.0 when operated in selected reaction monitoring mode (SRM) with Q1 and Q3 resolution set to 0.4 and 0.7 Da FWHM, respectively. | 
| Gas requirements | Collision gas: 99.995% pure argonCollision gas supply pressure: 135 ± 70 kPa (20 ± 10 psig)
 Sheath/aux/sweep gas: 99% pure nitrogen
 Sheath/aux/sweep gas supply pressure: 690 ± 140 kPa (100 ± 20 psig)
 Maximum sheath gas consumption: ≈20 L/min
 | 
| Environment requirements | Functional temperature range: 15–27 °C (59–81 °F)Optimal temperature range: 18–21 °C (65–70 °F)
 Heat output: 1550 W (5400 Btu/h)
 Total system heat output: 4420 W (15,380 Btu/h)
 Particulate matter: <3,500,000 particles per cubic meter of air (<100,000 particles of >5 µm diameter per cubic foot of air)
 Relative humidity: 20% to 80%, without condensation
 Floors must be free of vibration
 | 
| Power consumption | 2300VA | 
| Power rating | 230V~, 50/60Hz, 10.0A max | 
| Dimensions (W x D x H) | 30 x 33 x 27 in (76 x 84 x 68 cm) | 
| Weight | 275 lb (125 kg) | 
 
| Manufacturer | Thermo Fisher Scientific | 
| Device | Column Compartment H | 
| Part No. | VH-C10-A | 
| Serial No. | 6301071 | 
| Mfg. date | 2016-11-11 | 
| Power consumption | 326VA | 
| Power rating | 100-240V~, 50/60Hz, 4AT | 
| Dimensions (W x D x H) | 4.4 x 16.5 x 27.6 in (11.1 x 42 x 70 cm) | 
| Weight | 28.7 lb (13 kg) | 
 
| Manufacturer | Thermo Fisher Scientific | 
| Device | Split Sampler HT | 
| Part No. | VH-A10-A | 
| Serial No. | 8302664 | 
| Mfg. date | 2016-04-21 | 
| Power consumption | 550VA | 
| Power rating | 100-240V~, 50/60Hz, 5AT | 
| Dimensions (W x D x H) | 16.5 x 11.4 x 24.4 in (42 x 29 x 62 cm) | 
 
| Manufacturer | Thermo Fisher Scientific | 
| Device | Binary Pump H | 
| Part No. | VH-P10-A | 
| Serial No. | 8302653 | 
| Mfg. date | 2016-04-21 | 
| Power consumption | 550VA | 
| Power rating | 100-240V~, 50/60Hz, 4AT | 
 
| Manufacturer | Peak Scientific | 
| Model | NM32LA | 
| Serial No. | A16-02-112 | 
| Gas type | Nitrogen | 
| Integrated compressor | Yes | 
| Application types | LC-MS or LC-MS/MS | 
| Max gas flow | 32L/min | 
| Max output pressure | 100psi/6.9bar | 
| Pressure gauges | 1 | 
| Gas outlets fitting | ¼" BSP | 
| Start up time | 30 mins | 
| Heat output | 5729 BTU | 
| Max operating temperature | 30°C (86°F) | 
| Particles | <0.01µm | 
| Noise level | 54 dBA @ 1m | 
| Accreditations | CSA, FCC, CE | 
| Power rating | ~230V, 50/60Hz, 7A | 
| Power consumption | 1680 VA | 
| Dimensions (W x D x H) | 23.6 x 29.5 x 28.1 in (60 x 75 x 71.3 cm) | 
| Weight | 226.6 lb (103 kg) |