Beckman Coulter Biomek NXp Automated Liquid Handling Workstation
Successful Software Communication & Plate Movement
After connecting the NXp to the Biomek software, we performed the following tests:
- NXp initializes and communicates with Biomek software: PASSED
- Home all of the motors: PASSED
- Prime syringe pumps: PASSED
- Aspirate/dispense fluids: PASSED
- Load/unload tips: PASSED
- Lift/move microplates: PASSED
- Create/edit/remove programs: PASSED
- Beckman Coulter Biomek NXp Automated Liquid Handler
- MultiChannel 96 P20/200 Head with Gripper
- PC with software installed
- Monitor
- Mouse, keyboard
- Tip loader
- Heating block with controller
- Accuframe sensor
- Accuframe tool
- Metal reservoir with adapter plate
- Serial cable
- Power cables
Consumables are required for this system to be used in production. You will need to acquire the consumables separately. They can generally be purchased from the manufacturer or a third-party dealer.
Manufacturer
|
Beckman Coulter
|
Model
|
Biomek NXp
|
Catalog No.
|
A31841
|
Serial No.
|
A318410148
|
Environment
|
Indoor use only
|
Compressed Air Requirements
NX-MC: Requires compressed air NX-S8: Only Span-8 instruments that use a Microplate Shaking ALP require compressed air
|
Air Pressure: 60 - 100 psi of compressed air 414 - 690 kPa of compressed air Air Flow: 2.25 cfm @ 60 psi 0.064 m3/min @ 414 kPa
|
System Fluid Requirements
NX-S8: Requires system fluid
|
De-ionized or Distilled water System fluid ambient operating temperature is 59°–86°F (15°–30°C). System fluid should be degassed for 24 hours prior to use.
|
Weight
|
190 lbs. (86.2 kg)
|
Ambient Operating Temperature
|
41°-86°F (5°-30°C)
|
Humidity Restrictions
|
<85% (non-condensing) @ 86°F (30°C)
|
Altitude Restrictions
|
up to 6562 ft (2000 m)
|
Installation Category
|
Category II
|
Pollution Degree
|
2
|
Sound Pressure Level
|
Maximum sound pressure: 82 dB Maximum sound pressure at 1 meter away: 75 dB
|
Fuses
|
US: 250 VAC, 3.15 amp, 5x20 mm, Slow-blow, UL recognized, CSA certified Europe: 250 VAC, 3.15 amp, 5x20 mm, Slow-blow, CENELEC approved
|
Communications to Host
|
RS-232 port
|
Communications to Tip Loader NX-MC: Uses a Tip Loader
|
CAN
|
Communications to Active ALPs via the Biomex NXp Instrument
|
CAN
|
Power Requirements
|
100-240V, 4A, 50/60Hz
|
Overall Dimensions (WxDxH)
|
35 x 31 x 42"
|