Our Testing Process

1. Microfluidizer technician inspects electrical components, verifying proper functionality. Rewiring is done if needed.
2. Microfluidizer technician performs a detailed inspection on the high pressure fittings, and proper tightness are ensured.
3. The low pressure fittings are then inspected for proper seal.
4. Microfluidizer technician inspects hydraulic an pneumatic systems for damage. Then the unit is turned on, to make sure it runs smoothly.
5. Microfluidizer technician changes hydro oil if necessary.
6. The unit is then run/sterilized with denatured alcohol.
7. The unit is then cleaned and stainless steel is polished.
8. Upon purchase by customer, plunger seal(s) are replaced, to ensure a fresh seal when the customer gets it. 

 

Available Upgrades, Accessories, and Options

- Interaction Chamber
- Stainless Steel Reservoir
- Mechanical Pressure Gauge
- Power Cable
- Extended Warranty Available


 

Microfluidics M-110P Microfluidizer Processor Applications

  • Cell Disruption - This is used for breaking open cells to extract proteins, nucleic acids, and other intracellular materials.
  • Emulsification - Creation of stable emulsions by dispersing one liquid into another, typically with the goal of producing uniform droplets in applications like cosmetics, food, and pharmaceuticals.
  • Nanoparticle and Liposome Preparation - Formation of nanoparticles or liposomes for drug delivery, diagnostics, and other applications requiring fine particle sizes. 
  • Homogenization - Ensuring uniform distribution of particles in a liquid, often for biological, food, or chemical products. 
  • Dispersing Powders or Solids into Liquids - Efficiently dispersing powders or solid materials into liquids, which is crucial for making stable suspensions and slurries. 
  • Particle Size Reduction - Reducing the size of particles in a sample to achieve desired size distributions, which is essential for the uniformity and quality of products in various sectors. 
  • Vaccine and Antigen Production - Disrupting bacterial or viral cells for vaccine preparation, enhancing the delivery of antigens in therapeutic formulations. 
  • Gene Therapy Applications - Creating gene delivery vectors or working with nucleic acids (e.g., DNA or RNA) for generic engineering or therapy. 
  • Bioprocessing - Large-scale production of biopharmaceuticals, enzymes, and other biologics through the processing of cell cultures or bio-based products.

 


M110P Microfluidizer Specifications

Manufacturer

Microfluidics

Model No.

M110P-II

Maximum operating pressure up to 30,000 psi (2068 bar)
Minimum sample size 50 ml

Flow rate

up to 120 mL/min

Product temperature limit

158°F (70°C)

Noise level

80 dB TWA

Power rating

110V, 60Hz, 1Ph

Dimensions (W x D x H)

34 x 23 x 20 in

Weight

300 lb (136 kg)

 

Need Assistance? We're Happy to Help!

  1. Live chat with us by clicking on the icon at the lower right of your screen.
  2. Email us at cs@newlifescientific.com
  3. Call us at (567)-210-2838

 

Estimate Shipping

Microfluidics M110P Electric-Hydraulic Microfluidizer

Payment & Security
  • American Express
  • Diners Club
  • Discover
  • Google Pay
  • Mastercard
  • Shop Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

READY TO BUY OR SELL USED LAB EQUIPMENT?

We can help. In addition to selling lab equipment, we also purchase it.

Contact us below for immediate service.

Ask a Question or Request a quote!