 Manufacturer 
 | 
 Agilent Technologies, Inc 
 | 
| 
 Model 
 | 
 Synergy Neo2 
 | 
| 
 Product No. 
 | 
 N2SMA-SN 
 | 
| 
 Serial No. 
 | 
 22042012 
 | 
| 
 Mfg. date 
 | 
 2022-04-20 
 | 
| 
 Detection modes 
 | 
 UV-Vis absorbance Fluorescence intensity Luminescence Fluorescence polarization Time-resolved fluorescence Alpha 
 | 
| 
 Read methods 
 | 
 End point, kinetic, spectral scanning, well-area scanning 
 | 
| 
 Microplate types 
 | 
 6- to 1536-well plates 
 | 
| 
 Other labware supported 
 | 
 Petri and cell culture dishes Agilent BioTek Take3 microvolume plates 
 | 
| 
 Temperature control 
 | 
 4-Zone Incubation to 70°C with Condensation Control ±0.2°C at 37°C 
 | 
| 
 Shaking 
 | 
 Linear, orbital, double orbital 
 | 
| 
 Software 
 | 
 Agilent BioTek Gen6 data analysis software Agilent BioTek Gen5 Secure software enables 21 CFR Part 11 compliance (option) 
 | 
| 
 Automation 
 | 
 Agilent BioTek BioStack microplate stacker and third-party automation compatible Agilent BioTek BioSpa 8 automated incubator compatible Agilent BenchCel microplate handler compatible 
 | 
| 
 CO2 and O2 control (Option) 
 | 
 Range: 0-20% (CO2); 1-19% (O2) Control resolution: ±0.1% (CO2 and O2) Stability: ±0.2% at 5% CO2; ±0.2% at 1% O2 Models for CO2/O2, or CO2 only, are available 
 | 
| 
 Barcode reader 
 | 
 1D and 2D camera-based scanner 
 | 
| 
 Absorbance 
 | 
  | 
| 
 Light source 
 | 
 Xenon flash 
 | 
| 
 Detector 
 | 
 Photodiode 
 | 
| 
 Wavelength selection 
 | 
 Monochromator 
 | 
| 
 Wavelength range 
 | 
 230-999 nm, 1nm increments 
 | 
| 
 Monochromator bandwidth 
 | 
 <3 nm (230-285 nm), <6 nm (>285 nm) 
 | 
| 
 Dynamic range 
 | 
 0-0.4 OD 
 | 
| 
 Resolution 
 | 
 0.0001 OD 
 | 
| 
 Path length correction 
 | 
 Yes 
 | 
| 
 Monochromator wavelength accuracy 
 | 
 ±2 nm 
 | 
| 
 Monochromator wavelength repeatability 
 | 
 ±0.2 nm 
 | 
| 
 OD accuracy 
 | 
 <1% at 2.0 OD <3% at 3.0 OD 
 | 
| 
 OD linearity 
 | 
 <1% from 0-3.0 OD 
 | 
| 
 OD repeatability 
 | 
 <0.5% at 2.0 OD 
 | 
| 
 Stray light 
 | 
 0.03% at 230 nm 
 | 
| 
 Reading speed (kinetic) 
 | 
 96 wells: 6 s 
 | 
| 
 Fluorescence Intensity 
 | 
  | 
| 
 Light source 
 | 
 Xenon flash 
 | 
| 
 Detector 
 | 
 PMT 
 | 
| 
 Wavelength selection 
 | 
 Quad monochromators (top/bottom) Filters (top/bottom) 
 | 
| 
 Wavelength range 
 | 
 Monochromators: 250-700 nm (850 nm option) Filters: 200-700 nm (850 nm option) 
 | 
| 
 Monochromator bandwidth 
 | 
 Variable, from 3 to 50 nm in 1 nm increments 
 | 
| 
 Dynamic range 
 | 
 7 decades 
 | 
| 
 Sensitivity 
 | 
 Filters: Fluorescein 0.2 pM (4 amol/well, 20 μL, 384-well low-volume plate)—top Fluorescein 1 pM (10 amol/well, 10 μL, 1536-well plate)—top Fluorescein 1 pM (0.1 fmol/well, 100 μL, 384-well plate)—bottom 
Quad monochromator: Fluorescein 2 pM (40 amol/well, 20 μL, 384-well low-volume plate)—top Fluorescein 2.5 pM (0.25 fmol/well, 100 μL, 384-well plate)—bottom 
 | 
| 
 Reading speed (kinetic) 
 | 
 96 wells: 6 s 384 wells: 11 s 1536 wells: 25 s 
 | 
| 
 Luminescence 
 | 
  | 
| 
 Wavelength range 
 | 
 300-700 nm 
 | 
| 
 Dynamic range 
 | 
 >6 decades 
 | 
| 
 Sensitivity 
 | 
 5 amol ATP typical (384-well low-volume plate) 
 | 
| 
 Fluorescence Polarization 
 | 
  | 
| 
 Light source 
 | 
 Xenon flash 
 | 
| 
 Detector 
 | 
 PMT 
 | 
| 
 Wavelength selection 
 | 
 Filters 
 | 
| 
 Wavelength range 
 | 
 280-700 nm (850 nm option) 
 | 
| 
 Sensitivity 
 | 
 1 mP standard deviation at 1 nM fluorescein 
 | 
| 
 Time-Resolve Fluorescence 
 | 
  | 
| 
 Light source 
 | 
 Xenon flash or TRF laser (option) 
 | 
| 
 Detector 
 | 
 Dual PMT or single PMT (option) 
 | 
| 
 Wavelength selection 
 | 
 Filters (top/bottom) 
 | 
| 
 Sensitivity 
 | 
 With TRF laser: 5 fM (384-well low-volume plate) With Xenon flash lamp: 40 fM (384-well low-volume plate) 
 | 
| 
 Alpha Detection 
 | 
  | 
| 
 Light source 
 | 
 100 mW 680 nm laser 
 | 
| 
 Detector 
 | 
 PMT 
 | 
| 
 Wavelength selection 
 | 
 Filters (top) 
 | 
| 
 Sensitivity 
 | 
 100 amol LCK peptide (384-well low-volume plate) 
 | 
| 
 Read speed 
 | 
 96 wells: 30 s 384 wells: 1 min 50 s 1536 wells: 7 min 20 s 
 | 
| 
 Reagent Dispensers 
 | 
  | 
| 
 Supported detection modes 
 | 
 All modes 
 | 
| 
 Number 
 | 
 2 syringe pumps 
 | 
| 
 Supported labware 
 | 
 6- to 384-well plates, Petri dishes 
 | 
| 
 Dead volume 
 | 
 1.1 mL, with backflush 
 | 
| 
 Dispense volume 
 | 
 5-1,000 µL in 1 µL increments 
 | 
| 
 Dispense accuracy 
 | 
 ±1 µL or 2% 
 | 
| 
 Dispense precision 
 | 
 ≤2% at 50-200 µL 
 | 
| 
 Physical characteristics 
 | 
  | 
| 
 Power rating 
 | 
 24VDC 
 | 
| 
 Power consumption 
 | 
 250W max. 
 | 
| 
 Dimensions (W x D x H) 
 | 
 Without TRF laser: 15.4 x 20.7 x 16.1 in (39.2 x 52.5 x 40.8 cm) With TRF laser: 15.4 x 24.2 x 16.1 in (39.2 x 61.4 x 40.8 cm) 
 | 
| 
 Weight 
 | 
 Without TRF laser: 78 lb (35.4 kg) With TRF laser: 100 lb (45.3 kg) 
 |