- Molecular Devices SpectraMax 190 Microplate Reader Absorbance Spectrophotometer
- Dell Laptop with Softmax Pro v7.0.2 Software Installed
- Communication Cable
- USB Cable
- Laptop Charger
- Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Molecular Devices SpectraMax 190 microplate reader spectrophotometer with the following results:
- Reader connects to Softmax Pro software: PASSED
- Ran spectrum scan of holmium oxide standard; results were accurate to published spectrum: PASSED
- Plate load and unload: PASSED
- Heating: PASSED
Manufacturer
|
Molecular Devices
|
Model
|
SpectraMax 190
|
Serial No.
|
NNR05228
|
Photometric performance specifications
|
|
Wavelength range
|
190‒850 nm
|
Wavelength selection
|
Monochromator, tunable in 1 nm increments
|
Wavelength bandwidth
|
2 nm
|
Wavelength accuracy
|
< ±1.0 nm
|
Wavelength repeatability
|
± 0.2 nm
|
Photometric range
|
0‒4.0 OD
|
Photometric resolution
|
0.001 OD
|
Photometric accuracy
|
< ±0.006 OD ±1.0%, 0‒2.0 OD
|
Photometric precision
|
< ±0.003 OD ±1.0%, 0‒2.0 OD
|
PathCheck sensor measurement error
|
< 5% compared to cuvette reading
|
Stray light
|
≤ 0.05% @ 230 nm
|
Light source
|
Xenon flash lamp
|
Microplate read times
|
Endpoint: 12 seconds Kinetics: 9 second min. interval
|
Temperature specifications
|
|
Temperature range
|
Ambient +4‒45°C
|
Temperature uniformity (microplate)
|
±0.5°C at 37°C, well-to-well
|
Temperature stabilization time
|
30 minutes maximum upon initiation
|
Reading chambers
|
Isothermal when temperature regulation not enabled, <1°C
|
General specifications
|
|
Operating temperature
|
10‒40°C
|
Storage temperature
|
5‒40°C
|
Power rating
|
100-240V, 50-60Hz, 4A
|
Power consumption
|
<250 W
|
Dimensions (W x D x H)
|
22.8 x 15 x 8.6 in (58 x 38 x 22 cm)
|
Weight
|
30 lb (13.6 kg)
|