After your instrument arrives at our facility, it will be inspected and tested to verify it is free of malfunction. Unless stated otherwise, we are confident it will perform at manufacturer specifications. If you receive this instrument and it does not meet your required specifications, your purchase will be protected by our warranty.
Specifications
Manufacturer
|
Molecular Devices
|
Model
|
M2
|
Absorbance Photometric Performance
|
|
Wavelength range
|
200-1000 nm
|
Wavelength selection
|
Monochromator, tunable in 1 nm increments
|
Wavelength bandwidth
|
≤ 4.0 nm full width half maximum
|
Wavelength accuracy
|
±2.0 nm across wavelength range
|
Wavelength repeatability
|
±0.2 nm
|
Photometric range
|
0.000 to 4.000 OD
|
Photometric resolution
|
0.001 OD
|
Photometric accuracy (Microplate)
|
< ± 0.006 OD ±1.0%, 0-2 OD
|
Photometric accuracy (Cuvette)
|
< ± 0.005 OD ±1.0%, 0-2 OD
|
Photometric precision
|
< ± 0.003 OD ±1.0%, 0-2 OD
|
Baseline flatness
|
<0.001 OD
|
Stray light
|
≤ 0.05% at 230 nm
|
Fluorescence Photometric Performance
|
|
Dual Monochromators
|
1 nm increment selection
|
EX Wavelength range
|
250-850 nm
|
EM Wavelength range
|
360-850 nm
|
Wavelength bandwidth (EX, EM)
|
9 nm
|
Top read optimized sensitivity
|
5 pM (fluorescein) in 96 wells
|
Top-read Guaranteed sensitivity
|
15 pM (fluorescein) in 96 wells
|
Top-Resolved Fluorescence (secondary mode)
|
|
EX Wavelength range
|
250-850 nm
|
EM Wavelength range
|
360-850 nm
|
Data collection
|
50-1450 μsec. increments
|
Guaranteed sensitivity
|
7 pM Eu-chelate
|
Luminescence (secondary mode)
|
|
Wavelength range (M2)
|
360-850 nm
|
Guaranteed sensitivity
|
50 fM alkaline phosphatase
|
General photometric performance
|
|
Plate formats
|
6, 12, 24, 48, 96, 384 wells
|
Light source
|
Xenon flash lamp (1 joule/flash)
|
Detector
|
Photomultiplier tube (PMT)
|
Read time 96 wells: 384 wells:
|
Abs 18 sec., FI 15 sec. Abs 49 sec., FI 45 sec.
|
Shaker time
|
0 to 999 seconds
|
Temperature control
|
Ambient +4°C to 45°C
|
Temperature uniformity
|
<1°C at 37°C set point
|
Temperature accuracy
|
土1°C at 37°C set point
|
General specifications
|
|
Power consumption
|
< 125 watts
|
Power
|
100-240 Vac, 3 A, 50/60 Hz
|
Dimensions
|
23x16x9
|
Weight
|
35lbs
|