- Molecular Devices SpectraMax M2 Multimode Microplate Reader
- Dell Latitude Laptop with SoftMax Pro 6.2.2 Installed
- Laptop Power Supply/Charger
- USB to Serial Cable
- Serial Adapter Cable
- Top Read Adapter for 96 and 384 Well Plates
- Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Molecular Devices SpectraMax M2 multimode microplate reader with the following results:
- Powers on, initializes, and connect to control software: PASSED
- Ran spectrum scan of holmium oxide standard; results match published standards: PASSED
- Cuvette reading: PASSED
- Plate loading/unloading: PASSED
- Plate heating: PASSED
- Plate cooling: PASSED
- Plate shaking: PASSED
Manufacturer
|
Molecular Devices
|
Model
|
SpectraMax M2
|
Serial No.
|
D02109
|
Absorbance photometric performance
|
|
Wavelength selection
|
Monochromator, tunable in 1 nm increments
|
Wavelength range
|
200-1000 nm
|
Wavelength bandwidth
|
≤4.0 nm
|
Wavelength accuracy
|
±2.0 nm
|
Wavelength repeatability
|
±0.2 nm
|
Photometric range
|
0-4.0 OD
|
Photometric resolution
|
0.001 OD
|
Photometric accuracy (microplate)
|
<±0.006 OD ±1.0%, 0-2 OD
|
Photometric accuracy (cuvette)
|
<±0.005 OD ±1.0%, 0-2 OD
|
Photometric precision
|
<±0.003 OD ±1.0%, 0-2 OD
|
Baseline flatness
|
<0.001 OD
|
Stray light
|
<0.5% @ 230 nm
|
Fluorescence photometric performance
|
|
Dual monochromators
|
1 nm increment selection
|
EX Wavelength range
|
250-850 nm
|
EM Wavelength range
|
360-850 nm
|
Wavelength bandwidth (EX, EM)
|
9 nm
|
Top-read optimized sensitivity
|
5 pM (fluorescein) in 96 wells
|
Top-read guaranteed sensitivity
|
15 pM (fluorescein) in 96 wells
|
Time-Resolved Fluorescence (secondary mode)
|
|
EX Wavelength range
|
250-850 nm
|
EM Wavelength range
|
360-850 nm
|
Data collection
|
50-1450 µsec., 200 µsec. increments
|
Guaranteed sensitivity
|
7 pM Eu-chelate
|
Luminescence (secondary mode)
|
|
Wavelength range
|
360-850 nm
|
Guaranteed sensitivity
|
50 fM alkaline phosphatase
|
General photometric performance
|
|
Plate formats
|
6, 12, 24 , 48, 96, 384 wells
|
Light source
|
Xenon flash lamp (1 joule/flash)
|
Detector
|
Photomultiplier tube (PMT)
|
Read time
|
96 wells: Abs 18 sec., Fl 15 sec. 384 wells: Abs 49 sec., Fl 45 sec.
|
Shaker time
|
0 to 999 seconds
|
Temperature control
|
Ambient +4°C to 45°C
|
Temperature uniformity
|
<1°C at 37°C set point
|
Temperature accuracy
|
±1°C at 37°C set point
|
General specifications
|
|
Power rating
|
100-240V, 50-60Hz, 3.5A
|
Power consumption
|
<125 watts
|
Dimensions (W x D x H)
|
22.8 x 15 x 8.6 in (58 x 38 x 22 cm)
|
Weight
|
35 lb (15.9 kg)
|