- Tecan Infinite M Nano+ 200 Pro Microplate Reader (30050303)
- ThinkPad Laptop with Tecan i-control Installed
- Microsoft Excel 2010
- USB Cable
- Laptop Charger
- Power Cable
- 90-Day Warranty
 
After its initial disinfection and inspection, our technicians tested the Tecan Infinite M Nano+ 200 Pro microplate reader with the following results: 
 
- Powers on and initializes: PASSED
- On board plate load button: PASSED
- Reader connects to software: PASSED
- Multiple test assays performed with consistent results: PASSED
 
Note: It is possible that some of these specifications do not apply to this item, specifically. They are covering the entire Infinite M200 line, not just the M Nano+ configuration.
| 
 Manufacturer 
 | 
 Tecan 
 | 
| 
 Model 
 | 
 Infinite 200 Pro 
 | 
| 
 Configuration 
 | 
 Infinite M Nano+ 
 | 
| 
 Reference No. 
 | 
 30050303 
 | 
| 
 Serial No. 
 | 
 1906004632 
 | 
| 
 Mfg. date 
 | 
 20190612 
 | 
| 
 Power rating 
 | 
 100-120VAC / 220-240VAC, 50/60Hz, 150VA 
 | 
| 
 Dimensions (W x D x H) 
 | 
 17 x 18 x 9.5 in 
 | 
| 
 Light source 
 | 
 UV Xenon flashlamp 
 | 
| 
 Wavelength selection 
 | 
 Quad4 Monochromators system (2 excitation and 2 emission monochromators) 
 | 
| 
 Bandwidth 
 | 
 Ex: < 5 nm for  λ ≤ 315 nm and < 9 nm for λ > 315 nm; Em: < 20 nm 
 | 
| 
 Wavelength accuracy 
 | 
 Absorbance: < ± 0.5 nm for λ > 315 nm; < ± 0.3 nm for λ ≤ 315 nm Fluorescence: < ± 2 nm for λ > 315 nm; < ± 1 nm for λ ≤ 315 nm 
 | 
| 
 Wavelength reproducibility 
 | 
 Absorbance: < ± 0.5 nm for λ > 315 nm; < ± 0.3 nm for λ ≤ 315 nm Fluorescence: < ± 1 nm for λ > 315 nm; < ± 0.5 nm for λ ≤ 315 nm 
 | 
| 
 Wavelength range 
 | 
 Fluorescence intensity: Ex 230 – 850 nm, Em 280 – 850 nm Absorbance: 230 - 1000 nm 
 | 
| 
 Detectors 
 | 
 Fluorescence – PMT, UV and red-sensitive Absorbance – UV silicon photodiode Luminescence –photon counting system with low dark current PMT 
 | 
| 
 Plate formats 
 | 
 6- to 384-well plates, cuvettes, NanoQuant plate 
 | 
| 
 Temperature control 
 | 
 Ambient +5°C up to 42°C 
 | 
| 
 Shaking 
 | 
 Linear, orbital 
 | 
| 
 Fluorescence sensitivity values 
 | 
 Top reading: 170 amol / well (100 μl; 384-well plate) Bottom reading: 9.0 fmol / well (200 μl; 96-well plate) TRF: 90 amol / well (100 μl; 384-well plate) 
 | 
| 
 Absorbance ratio accuracy @ 260 / 280 nm 
 | 
 ± 0.07 
 | 
| 
 Precision @ 260 nm 
 | 
 < 0.2% 
 | 
| 
 Accuracy @ 260 nm 
 | 
 < 0.5% 
 | 
| 
 Measurement range 
 | 
 0 - 4 OD 
 | 
| 
 Injector pump speed 
 | 
 100 - 300µl/s 
 | 
| 
 Injection volume 
 | 
 selectable in 1µl increments; max. volume: 800 µl per stroke 
 | 
| 
 Dead volume 
 | 
 100 µl including pump back 
 | 
| 
 Fastest read times 
 | 
 96-well plate: 20 sec 384-well plate: 30 sec Wavelength Ex / Em-scan, 96-well plate (450 - 550 nm, 5 nm step): 150 sec 
 |