- New Brunswick Scientific Innova 4230 Refrigerated Incubator Shaker (121408-001)
- Shaker Platform
- Shelf
- Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the New Brunswick Scientific Innova 4230 refrigerated incubator shaker with the following results:
- Shaking holds steady at setpoint: PASSED
- Shaking produces no excessive noise or vibration: PASSED
- Shaker calibration verification: PASSED
- Heating holds steady at 37°C setpoint: PASSED
- Cooling holds steady at 4.0°C setpoint: PASSED
- Temperature reading calibration: PASSED
- Installed a new belt.
- Installed new feet.
|
Manufacturer
|
New Brunswick Scientific
|
|
Model
|
Innova 4230
|
|
Mfg. No.
|
121408-001
|
|
Serial No.
|
200233303
|
|
Mfg. date
|
2002
|
|
Shaking range
|
25 to 400 rpm (25 to 300 rpm for stacked units)
|
|
Temperature range
|
20°C below ambient temperature with minimum of 4.0°C
+75°C maximum for 50Hz units or 80°C for 60Hz units
|
|
Shelf dimensions (W x D)
|
20 x 20 in (50.8 x 50.8 cm)
|
|
Power rating
|
120V, 50/60Hz, 1Ph
|
|
Power consumption
|
1800VA
|
|
Platform dimensions (W x D)
|
18 x 18 in (45.7 x 45.7 cm)
|
|
Overall dimensions (W x D x H)
|
24 x 28 x 37 in (61 x 71 x 94 cm)
|
|
Weight
|
360 lb (163 kg)
|