- Perkin Elmer Victor X2 Multimode Plate Reader
- PC with Perkin Elmer Interface Card, Monitor, Keyboard, and Mouse
- All necessary Data Cables
- All necessary Power Cables
- 90-day Warranty
After its initial disinfection and inspection, our technicians tested the Perkin Elmer Victor X2 multimode plate reader with the following results:
- Generated OS on the PC hard drive and installed software: PASSED
- Reader and PC initialize and connect: PASSED
- Ran several luminescence scans with appropriate results: PASSED
- Ran fluorescence scan which resulted in a filter slide error: FAILED*
*Attempting a fluorescence scan results in the following error message: “EMISSION FILTER SLIDE ERROR AT MEASUREMENT UNIT. The slide cannot move freely, the EEPROM parameters are corrupted or there is a mechanical/electronic fault in the slide mover unit.”
NOTICE! While this instrument is covered by a 90-day warranty, this warranty applies to everything except for the functionality of the fluorescence read mode.
Manufacturer
|
Perkin Elmer
|
Model
|
Victor X2
|
Item No.
|
2030-0020
|
Serial No.
|
20300123
|
Light Sources
|
Tungsten-halogen lamp
|
Detection Units
|
Photomultiplier tube
|
Plate Formats
|
1 to 1536
|
Shaking
|
Linear, orbital and double orbital. Duration, speed and amplitude are adjustable.
|
Temperature Control
|
Temperature range from 2°C above ambient up to 50°C (0.1°C increments)
|
Uniformity
|
+/-1°C
|
Heating Time
|
<10 minutes (from RT up to 37°C)
|
Dispense Volume
|
5-350uL (1uL increments)
|
Typical Accuracy
|
<5% for 5uL, <0.5% for 50uL, <0.05% for 350uL
|
Typical Precision
|
<1.4% for 5uL, <0.2% for 50uL, <0.02% for 350uL
|
Dead Volume
|
<0.5mL
|
Electrical requirements
|
110-120 V/220-240V, 50/60Hz
|
Luminometry 96-Well Plate
|
Lower Limit of Detection (LLD)
|
Total flux of 20,000 photons/s (standard PMT) 100,000 photons/s (red sensitive PMT)
|
ALP
|
1 amol/well with AMPPD substrate
|
Flash Luminescence
|
80 amol/well ATP
|
Glow Luminescence
|
0.9 pg/well using britelite
|
Fluorometry 96-Well Black Plate (200uL)
|
Fluorescein
|
<2 fmol/well, 10 pM
|
Linearity
|
>5 decades
|
Crosstalk
|
<0.01%
|
Umbelliferone
|
<200 fmol/well, 1 nM
|
Rhodamine
|
<100 fmol/well, 0.5 nM
|
Physical Data
|
Dimensions (W x D x H)
|
20 x 24 x 15 in
|
Weight
|
101 lbs.
|