- Perkin Elmer Victor X3 Multimode Microplate Reader (2030-0030)
- Load and Unload Microplate Stacks
- Dell Latitude Laptop with PerkinElmer 2030 Workstation Software Installed
- USB 2.0 to ARCNET Adapter
- USB Cable
- Coaxial Cable
- Laptop Charger
- x2 Power Cables
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Perkin Elmer Victor X3 multimode microplate reader with the following results:
- Reader initializes and connects to control software: PASSED
- Ran multiple scan; results were appropriate and repeatable: PASSED
- Plate stack loading and unloading: PASSED
- Fluorescence intensity: top and bottom
- Luminescence: flash and glow
- Absorbance (UV and visible)
Manufacturer
|
Perkin Elmer
|
Model
|
Victor X3
|
Model No.
|
2030-0030
|
Serial No.
|
20301912
|
Detection methods
|
Luminescence (flash and glow) Fluorescence Intensity (top and bottom reading) Absorbance (UV and visible)
|
Software
|
Perkin Elmer 2030 Manager v.4
|
Luminescence Detection (96-well plate)
|
Lower limit of detection (LLD)
|
Total flux of 20,000 photos/s (standard PMT) 100,000 photon/s (red-sensitive PMT)
|
ALP
|
1 amol/well with AMPPD substrate
|
Flash luminescence
|
80 amol/well ATP
|
Glow luminescence
|
0.9 pg/well using britelite
|
Absorbance Detection (96-well plate)
|
Measuring range @ 405 nm
|
0-4 A
|
Accuracy @ 405 nm
|
<2% (or 0.01 A) within 0-2 A
|
Precision @ 405 nm
|
<0.5% (or 0.01 A) within 0-2 A
|
Fluorescence Detection (200 uL, 96-well back plate)
|
Fluorescein
|
<2 fmol/well, 10 pM
|
Linearity
|
>5 decades
|
Crosstalk
|
<0.01%
|
Umbelliferone
|
<200 fmol/well, 1 nM
|
Rhodamine
|
<100 fmol/well, 0.5 nM
|
Other Specifications
|
Power rating
|
110-120V, 50/60Hz 220-240V, 50/60Hz
|
Power consumption
|
400VA
|
Dimensions (W x D x H)
|
20 x 24 x 15 in
|
Weight
|
101.4 lb (46 kg)
|