Manufacturer
|
Perkin Elmer
|
Model
|
Victor 3
|
Item No.
|
1420-012
|
Serial No.
|
4205843
|
Plate formats
|
1 to 1536
|
Shaking
|
Linear, orbital, and double orbital Duration, speed and amplitude are adjustable
|
Temperature Control
|
Temperature range
|
from 2°C above ambient up to 45°C (0.1°C increments)
|
Uniformity
|
±1°C
|
Heating time
|
<10 minutes (from RT up to 37°C
|
Luminescence Detection (96-Well Plate)
|
Lower Limit of Detection (LLD)
|
Total flux of 20,000 photons/s (standard PMT) 100,000 photons/s (red-sensitive PMT)
|
ALP
|
typically 1 amol/well with AMPPD substrate
|
ATP
|
Typically 80 amol/well in flash assays
|
Luciferase
|
0.9 pg/well using steadylite
|
Fluorescence Polarization (384-well black plate)
|
Fluorescein
|
1 nM, 40 µL
|
Standard deviation
|
<5 mP
|
Absorbance Detection (96-well plate)
|
Measuring range @ 405 nm
|
0-4 A
|
Accuracy @ 405 nm
|
<2% (or 0.01 A) within 0-2 A
|
Precision @ 405 nm
|
<0.5% (or 0.01 A) within 0-2 A
|
TR-Fluorescence Detection (200 µL,96-well clear plate)
|
Europium
|
Typically <10 amol/well, 50 fM
|
Linearity
|
>5 decades
|
Crosstalk
|
<0.01%
|
Terbium
|
typically <5 amol/well, 25 fM
|
Samarium
|
typically <50 amol/well, 250 fM
|
Dysprosium
|
typically <150 amol/well, 750 fM
|
Fluorescence Detection (200 µL, 96-well black plate)
|
Fluorescein
|
typically <2 fmol/well, 10 pM
|
Linearity
|
>5 decades
|
Crosstalk
|
<0.01%
|
Umbelliferone
|
typically <200 fmol/well, 1 nM
|
Rhodamine
|
typically <100 fmol/well, 0.5 nM
|
Physical Specifications
|
Main voltage
|
110-120V/220-240V, 50/60Hz
|
Power consumption
|
250VA
|
Dimensions (W x D x H)
|
19 x 23.5 x 15.5 in (485 x 590 x 383 mm)
|
Weight
|
108 lb (49 kg)
|