Fully Tested and Working Without Issue
After its initial disinfection and inspection, our technicians tested this epMotion 5070 with the following results:
- Initializes without error: PASSED
- Connects to controller: PASSED
- Executes programs: PASSED
- Robotic arm is smooth: PASSED
- Sensors working: PASSED
- Aspiration works: PASSED
- Tip disposal works: PASSED
- Eppendorf epMotion 5070 Robotic Pipette Liquid Handler
- Controller
- Mouse
- 0-50uL Single-Channel Tool
- Open box of 50uL Pipette Tips
- Power and comms cables
Manufacturer
|
Eppendorf
|
Model
|
epMotion 5070
|
Serial No.
|
5070 01009
|
Operation
|
Electronic
|
Pipetting type
|
Air-cushion
|
SLAS/ANSI positions
|
4
|
Application
|
PCR/qPCR and normalization
|
Volume range
|
0.2 uL - 1 mL
|
Power supply
|
100-240V +/-10%, 50-60Hz +/-5%
|
Max. power consumption
|
150W
|
Dimensions (W x D x H)
|
26 x 19 x 25 in. (65 x 48 x 63 cm)
|
Weight w/o accessories
|
99.2 lbs (45 kg)
|
Optical sensor
|
Contact-free recognition of fill level, inserted tools, labware, tip types and quantities
|
Conductor — X, Y, Z positioning
|
Random measurement error: +/-0.1 mm Systematic measurement error: +/-0.3 mm
|
Dispensing tool
Note: All measurements taken in pipetting mode, free jet, without pre-wetting, with distilled water, at 20°C
|
Random measurement error (1 µL)
|
≤3 %
|
Random measurement error (1,000 µL)
|
≤0.4 %
|
Systematic measurement error (1 µL)
|
±5 %
|
Systematic measurement error (1,000 µL)1,2
|
±0.7 %
|
Systematic measurement error (50 µL)1,2
|
±1.2 %
|