- Sakura Tissue-Tek TEC 6-EM-A1 Tissue Embedding Module (5108)
- Sakura Tissue-Tek TEC 6-CM-A1 Cryo Module (5109)
- Communication Cable
- x2 Power Cables
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Sakura Tissue-Tek TEC 6 tissue embedding module and cryo module with the following results:
- Touchscreen: PASSED
- Mold warmer: PASSED
- Sample warmer: PASSED
- Work surface heat: PASSED
- Work surface light: PASSED
- Paraffin tank: PASSED
- Cold spot: PASSED
- Paraffin dispenser: PASSED
- Cryo module: PASSED
Notice: This embedding center and cold plate have some cosmetic blemishes. Please review the available photos to see the physical condition before purchasing.
|
Manufacturer
|
Sakura Finetek
|
|
Product line
|
Tissue-Tek TEC 6
|
|
Product type
|
Embedding module
|
|
Product code
|
5108
|
|
Model
|
TEC6-EM-A1
|
|
Serial No.
|
51080467-0820
|
|
Mfg. date
|
August 2020
|
|
Temperature control
|
Paraffin chamber: 50 to 75°C Left and right warming chambers: 50 to 75°C Hot plate: 50 to 75°C
|
|
Cold spot temperature
|
15°C or below (room temperature of 25°C, no wind, no load (nothing on the cold spot))
|
|
Safety device
|
Overcurrent protection: Circuit protector Heater overheat protection: Self-holding over temperature protector, temperature fuse)
|
|
Location of the instrument
|
Indoor use
|
|
Power rating
|
115V, 60Hz, 9.0A
|
|
Dimensions (W x D x H)
|
23 x 25 x 15 in (58 x 64 x 38 cm)
|
|
Weight
|
Approximately 57 lb (26 kg)
|
|
Manufacturer
|
Sakura Finetek
|
|
Product line
|
Tissue-Tek TEC 6
|
|
Product type
|
Cryo module
|
|
Product code
|
5109
|
|
Model
|
TEC6-CM-A1
|
|
Serial No.
|
51093140-1024
|
|
Mfg. date
|
October 2024
|
|
Temperature range
|
-10°C to 0°C in 1°C increments
|
|
Lowest temperature achievable
|
-5°C or below (Room temperature of 25°C, with no wind, under no loaded environment)
|
|
Time to reach -5°C
|
Within 30 min after started to operate (Room temperature of 25°C, with no wind, measured at a center of the cooling plate under no loaded environment)
|
|
Plate surface capacity
|
60 cassette specimens
|
|
Safety device
|
Overcurrent protection (circuit protector) Overload protection for compressor (overload protection relay) Detection of fan rotation (built-in fan sensor)
|
|
Location of the instrument
|
Indoor use
|
|
Plate dimensions (W x D)
|
11.8 x 14.5 in (30 x 37 cm)
|
|
Power rating
|
115V, 60Hz, 2A
|
|
Dimensions (W x D x H)
|
13 x 25 x 15 in (33 x 62 x 38 cm)
|
|
Weight
|
Approximately 48 lb (22 kg)
|