- GeneVac HT-24 Centrifugal Vacuum Evaporator System (dual HT-12)
- x24 GeneVac Rotor Swing Buckets
- x2 GeneVac Condensers
- Waste Container
- Boc Edwards XDS10 Dry Scroll Vacuum Pump with Vacuum Hose
- Power Cables
- 120-Day Warranty
After its initial disinfection and inspection, our technicians tested the GeneVac HT-24 centrifugal vacuum evaporator system with the following results:
- Chamber heating: PASSED
- Chamber cooling: PASSED
- Rotor speed: PASSED
- Sample temperature: PASSED
- Condenser cooling: PASSED
- Condenser temperature: PASSED
- Condenser defrost: PASSED
- Vacuum integrity: PASSED
- Door open and close: PASSED
- Replaced motor drive belts.
Manufacturer
|
GeneVac
|
Product
|
Evaporator
|
Model
|
HT12 Series II
|
Serial No.
|
6879 (HT12) 6881 (HT12) T1048 (HT24)
|
Mfg. date
|
December 2010
|
Max rotor speed
|
1300 low speed 1750 high speed 1800 Dri-Pure
|
Max force
|
300-500G
|
Drive system
|
Direct
|
Operation imbalance
|
85g
|
Max load
|
12 x 1.5kg @ 500G
|
IR lamps number
|
6
|
Weight
|
180 kg
|
Condenser temperature
|
Minimum: -45°C Maximum: +60°C
|
Vacuum condenser capacity
|
3.0 liters
|
Exhaust condenser capacity
|
1.5 liters
|
Condenser level detector
|
No
|
Rapid defrost
|
Yes
|
Condenser chamber
|
316 Stainless steel
|
Condenser drain valve
|
Stainless steel/PTFE
|
Vacuum system pressure display/resolution
|
0 - 1000 mbar
|
Vacuum control
|
0 - 1000 mbar
|
Dri-Pure
|
Yes
|
Auto vacuum vent valve
|
Yes
|
Ultimate system vacuum
|
0.4 mbar
|
Vacuum pump - Scroll
|
Weight: 28 kg Maximum vacuum: 0.15 mbar (50Hz)/0.12mbar (60Hz) Flow rate: 3.6 m^3h^-1
|
Vacuum pump - CVP
|
Weight: 52 kg Maximum vacuum: 0.15 mbar Flow rate: 3.6 m^3h^-1
|
Power rating
|
208V/60Hz, single phase 9.1A
|
Dimensions (W x D x H)
|
22 x 27 x 33.5 in (HT12 single unit) 53 x 28 x 33.75 in (HT24 cabinet) 53 x 28 x 68 in (overall system)
|
Manufacturer
|
Boc Edwards
|
Product
|
Vacuum pump
|
Model
|
XDS10
|
Code No.
|
A726-01-903
|
Serial No.
|
041456436
|
Mass
|