Overview of SP Scientific VirTis 50L Ultra Cleanroom Freeze Dryer Lyophilizer

This is a cleanroom configuration VirTis Ultra 50L lyophilizer. A cleanroom configuration isolates the shelves completely away from all the running components of the unit from any components that would contaminate your samples in the clean room in front. 

The Ultra 50L lyophilizer has a maximum condenser capacity of 50 liters. Its lowest shelf temperature reaches to -67 degrees Celsius (50Hz) or -70 degrees Celsius (60Hz). While its lowest condenser temperature reaches to -82 degrees Celsius (50Hz) or -85 degrees Celsius (60Hz).

View in stock used freeze dryers/lyophilizers for sale here. Need assistance? Get in touch with us today.

 

VirTis 50L Ultra Cleanroom Features

 Steam-in-place (SIP) Feature - Ensures sterility without manual cleaning.

 Clean-in-place (CIP) Feature - Minimizes contamination risk.

 Smooth interior design prevents particle accumulation.

 Low particle emissions maintain ISO class rating.

 Isolator and barrier systems enhance sterility assurance.

 SCADA-based control systems enable audit trails and batch records.



SP VirTis 50L Ultra Freeze Dryer Lyophilizer Specifications

Manufacturer

SP Scientific

Model

50L Ultra EL-85°

Lowest shelf temperature (50Hz / 60Hz)

≤ -67°C / -70°C

Shelf temperature control range

-55 to 65°C

Shelf pull-down from 20°C to -40°C

≤ 40 minutes

Lowest condenser temperature (50Hz / 60 Hz)

≤ -82°C / -85°C

Maximum condenser capacity

≥ 50 L

Condenser surface area

10 ft2 (.93 m2)

Condenser pull-down from 20°C to -45°C

≤ 35 minutes

Maximum ice condensing capacity in 24 hours

≥ 20 L

Maximum deposition rate

≥ 0.83 L/hour

Number of compressors

2

Compressor horsepower

3.5 hp, 1.5 hp

Vacuum time to 100 mTorr

≤ 45 minutes

Vacuum rate of rise

≤ 30 mT/hour

Vacuum control +/- 5 mTorr

50mTorr to 500mTorr

Volume-based leak rate

≤ .0042 mbar L/sec

Lowest system vacuum

≤ 15 mT

Temperature uniformity

± 1.0°C

Construction

316L stainless steel shelves, product chamber and condenser chamber

Stoppering

Bottom-up hydraulic

Defrost type

Hot gas

Refrigerant type

CFC-free

Vapor port

8 in (20.3 cm)

Power rating

208-230VAC, 60Hz, 50A, 1Ph

220VAC, 50Hz

Recommended outlet

NEMA 6-50R

Overall dimensions (W x D x H)

25 x 78 x 75.75 in (64 x 198 x 192 cm)

Minimum clearance

10 in (25.4 cm)

Weight

2,000 lb (909 kg)

Note: Performance specifications are based on SP test data from units operating at an ambient room temperature of approximately 20°C. SP recommends an optimum operation range of 15-25°C (59-77°F).

Note: SP recommends a 24-inch (61 cm) clearance around all sides of the unit for serviceability. If machines are placed side by side, increase the minimum clearance to 48 inches (121.9 cm)





Our Testing Process

1. Freeze dryer is visually inspected on arrival and then a performance test is initiated to see how well the unit is operating.

2. The interior and exterior of the unit are cleaned using a dry ice blasting method to remove buildup and contaminants.

3. The freeze dryer technician checks the ambient pressures of the high and low side compressors (if applicable). A functional test of the refrigeration circuit is performed to verify that the unit is cooling in accordance with the manufacturer's specifications.

4. Air-cooled units use a condenser and fans to remove heat from the system. The condenser is inspected and cleaned to ensure proper airflow. Water-cooled units use facility water or a standalone chiller to dissipate heat from the refrigeration circuit. The water circuit is inspected for leaks and verified to be functioning correctly.

5. For rotary vane vacuum pumps, the oil is flushed and refilled. Scroll vacuum pumps, which are oil-free, are inspected and calibrated accordingly.

6. For benchtop and floor model freeze dryers equipped with tree and drum manifolds, the valves are cleaned, inspected, and regreased to ensure proper sealing and functionality.

7. The unit’s temperature sensors are calibrated to ensure accurate readings and proper performance (on non-Lyo software units).

8. Door seals and other critical components are inspected and replaced as needed.

9. Refrigeration lines are reinsulated, and all vacuum hoses are inspected or replaced to ensure proper system integrity.

10. If the unit includes software, the technician runs functionality tests, including checking the minimum and maximum achievable temperatures.

11. The freeze dryer technician verifies that the stoppering mechanism (if present) and isolation valves are functioning correctly.

Note: Isolation valves are designed to prevent air leaks within the system and must be tested for proper sealing during operation, heating, and defrost cycles.

12. The unit is repainted and redecaled where necessary to ensure a professional appearance.



 

NLS' Standard SP VirTis 50L Ultra Lyophilizer Accessories

 Vacuum Pump
 Shelves
 PC and LyoS 2.0 or Wizard software installed
 Corrosive-resistant Glass Door
  Power cable
  90-Day to 1-Year Warranty

 

Our Freeze Dryer Technician Checklist

At New Life Scientific, our technicians test each device thoroughly before shipping to verify performance and functionality. After our freeze dryer technician finishes the initial testing process, this Freeze Dryer Technician Checklist is filled out to record the important information gathered throughout the repair process. This checklist will be available publicly after the unit has undergone tech completion and is ready for the customers.

 

SP VirTis 50L Ultra Freeze Dryer Lyophilizer Applications

Biopharmaceutical Labs - Labs involved in the production of biologics, including monoclonal antibodies, therapeutic proteins, and peptides, use freeze dryers to stabilize and preserve these sensitive products.

Vaccine Development Labs - Labs working on vaccine production, especially those requiring lyophilization to preserve the potency of live or inactivated viruses, often rely on freeze-drying.

Gene and Cell Therapy Labs - Labs focused on gene therapies (e.g., plasmids, mRNA) and cell therapy (e.g., stem cells) use freeze dryers to preserve the viability and stability of biological materials.

Diagnostic and Assay Labs - Laboratories producing diagnostic kits, reagents, and assays, especially for long-term storage and stability, use freeze dryers to maintain their integrity over time.

Research and Biotech Labs - Labs involved in biological research, including the preservation of biological samples, cell lines, and microbial cultures, use freeze dryers for sample storage and transportation.

Blood Bank and Plasma Labs - Labs that process and store blood plasma, serum, and other biological materials often use freeze dryers to maintain the stability of these sensitive components.

Nutraceutical and Exosome Therapy Labs - Labs producing nutraceuticals (e.g., probiotics) or exosome-based therapies use freeze dryers to stabilize these products for longer shelf life.

 

Contact New Life Scientific Today for Additional Information

Our team can help you decide whether the VirTis 50L Ultra Cleanroom’s capacity and features align with your processing needs. Contact us today if you have questions or need additional specifications. You may also:

Live chat with us by clicking on the icon at the lower right of your screen.

Email us at cs@newlifescientific.com

Call us at (567)-210-2838

Estimate Shipping

SP Scientific VirTis 50L Ultra Cleanroom Freeze Dryer Lyophilizer

Payment & Security
  • American Express
  • Diners Club
  • Discover
  • Google Pay
  • Mastercard
  • Shop Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

READY TO BUY OR SELL USED LAB EQUIPMENT?

We can help. In addition to selling lab equipment, we also purchase it.

Contact us below for immediate service.

Ask a Question or Request a quote!