- Kendro Sorvall Discovery 100SE Floor Ultracentrifuge
After its initial disinfection and inspection, our technicians tested the Kendro Sorvall Discovery 100SE Floor Ultracentrifuge with the following results:
- Ran to 90,000 rpm: PASSED
- Cooling held at 5.6ºC: PASSED
- Performed inspection; no problem found: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Requires 208V electrical receptacle
Manufacturer
|
Kendro Sorvall
|
Model
|
Discovery 100 SE
|
Serial Number
|
Y10O-156697-YO
|
Mfg. Date
|
2003
|
Max Speed
|
100,000 RPM
|
Max RCF
|
802,006 x g
|
Speed control accuracy
|
+/- 10 RPM (1,000 RPM to max speed)
|
Time for acceleration and deceleration
|
About 5 minutes from 0 to 100,000 RPM or vice versa
|
Rotor temperature control/display accuracy
|
+/-0.5ºC (set temperature is from 0ºC to 40ºC)
|
Set speed
|
1,000 RPM to maximum speed in increments of 100 RPM
|
Vacuum system
|
Oil rotary vacuum pump and oil diffusion pump combined. Ultimate vacuum of 0.13 Pa
|
Interface
|
RS-232C x 1 CH
|
Weight
|
882 LBS. (400 kg)
|
Power consumption
|
3.1 Kw
|
Electrical Requirements
|
208V, 50/60Hz, 20A max
|
Dimensions (WxDxH)
|
31 x 27 x 39 in.
|