- Lauda Integral T 1200W Process Thermostat Recirculating Chiller 208-230V (LWP801)
- Attached 230V Power Cable
- 120-Day Warranty
After its initial disinfection and inspection, our technicians tested the Lauda Integral T 1200W recirculating chiller with the following results:
- Cooling temperature of 5°C: PASSED
- Heating temperature of 60°C: PASSED
- User interface: PASSED
Manufacturer
|
Lauda
|
Type
|
T 1200
|
Catalog No.
|
LWP 801
|
Serial No.
|
LWP801-11-0042
|
Working temperature min.
|
-25°C
|
Working temperature max.
|
120°C
|
Ambient temperature min.
|
5°C
|
Ambient temperature max.
|
40°C
|
Temperature stability
|
0.2±K
|
Application
|
external
|
Filing volume min.
|
3 L
|
Filing volume max.
|
7 L
|
Heater power
|
2.25kW
|
Cooling output at -25°C measured with ethanol
|
0.1kW
|
Cooling output at -20°C measured with ethanol
|
0.25kW
|
Cooling output at -15°C measured with ethanol
|
0.45kW
|
Cooling output at -10°C measured with ethanol
|
0.7kW
|
Cooling output at -5°C measured with ethanol
|
0.85kW
|
Cooling output at 0°C measured with ethanol
|
1.1kW
|
Cooling output at 5°C measured with ethanol
|
1.2kW
|
Cooling output at 10°C measured with ethanol
|
1.3kW
|
Cooling output at 20°C measured with ethanol
|
1.6kW
|
Interface
|
RS-232/485
|
Tube size
|
0.75”
|
Pump connection thread
|
G ¾ /(15)
|
Pump pressure max.
|
3.2 bar
|
Pump flow max. (pressure)
|
40 L/min
|
Refrigerant
|
R404A
|
Filing charge
|
1.1 kg
|
PS high pressure
|
28 bar
|
PS low pressure
|
19 bar
|
Noise level
|
58 dB(A)
|
Power rating
|
208-230V, 60Hz
|
Power consumption
|
2.7kW
|
Dimensions (W x D x H)
|
17.7 x 21.6 x 31.1 in (45 x 55 x 79 cm)
|