- Agilent BioTek 405 TS Microplate Washer (405TSRVS-SN)
- 96-tube Single Manifold Installed
- Agilent Vacuum Pump 3 (7103035)
- x4 Supply (Buffer) Bottles
- x2 Waste Bottles with Waste Sensor
- Tubing
- x3 Bottle Racks
- x2 Power Cables
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Agilent BioTek 405 TS microplate washer with the following results:
- Ran several wash cycles: PASSED
- Priming: PASSED
- Dispensing: PASSED
- Aspiration: PASSED
- Performed inspection; no issue found: PASSED
|
Manufacturer
|
Agilent Technologies, Inc
|
|
Model
|
405 TS
|
|
Product No.
|
405TSRVS-SN
|
|
Regulatory Model No.
|
405TSRVS
|
|
Serial No.
|
23050902
|
|
Mfg. date
|
2023-05-09
|
|
Includes
|
Automated Internal 4-Buffer Switching; Ultrasonic Advantage; 96-well microplate washing
|
|
Assays
|
ELISA, Cell-based assays (model dependent), Magnetic and polystyrene bead, Multiplex assays, Bead-based ELISAs, Filtration-to-waste processes
|
|
Microplate type
|
96-well standard, solid and filter-bottom
|
|
Manifold type
|
Single manifold with 96 sets of aspirate and dispense tubes arranged in an 8x12 array. Single manifolds can only process 96-well microplates.
|
|
Average residual volume (evacuation efficiency)
|
Average residual volume in the microwells is ≤ 2 µL per well after a 3-cycle wash, when 300 µL of deionized water with 0.1% Tween 20®, or buffer equivalent, is dispensed per well into a Costar® 96-well flat-bottom plate. The aspirate height adjustment is optimized for the plate prior to testing.
|
|
Dispense precision
|
≤ 3.0% CV when dispensing 300 µL per well of deionized water with 0.1% Tween 20, with FD&C #1 blue dye at rate 6 into a Costar 96-well flat-bottomed plate. The absorbance of the solution is read at 630 nm and 450 nm reference.
|
|
User interface
|
5.7” touch screen
|
|
Accessory outlet
|
≤5.0A, use for vacuum pump
|
|
Power rating
|
100-240V, 50/60Hz, 9.0A
|
|
Dimensions (W x D x H)
|
14 x 17 x 10 in (35.6 x 43.2 x 25.4 cm)
|
|
Weight
|
36 lb (16.5 kg)
|