- Barnstead Lab-Line MaxQ Mini 4000 Analog Benchtop Orbital Incubator Shaker (SHKA4450)
- Shaker Platform
- Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Barnstead Lab-Line MaxQ Mini 4000 analog benchtop incubator shaker with the following results:
- Shaking with no excessive noise or vibration: PASSED
- Shaker calibration: PASSED
- Heating to 37°C setpoint: PASSED
- Temperature calibration: PASSED
- Installed new gas springs.
- Installed a new belt.
Notice: This shaker has some minor cosmetic blemishes. Please review the available photos to see the physical condition before purchasing.
|
Manufacturer
|
Barnstead Lab-Line
|
|
Model
|
MaxQ Mini 4000
|
|
Model No.
|
SHKA4450
|
|
MO#
|
462140
|
|
Serial No.
|
0404-4872
|
|
Speed range
|
40-400 rpm
|
|
Temperature range
|
5° above ambient to 80°C
|
|
Temperature accuracy
|
±0.1°C at 37°C in flask
|
|
Maximum load
|
35 lb (15.9 kg)
|
|
Timer
|
Continuous or timed operation from 1 minute to 60 minutes
|
|
Display
|
LED display indicates temperature in 0.1°C increments. Analog tachometer displays speed in rpm
|
|
Orbit diameter
|
0.75 in (1.9 cm)
|
|
Mutable alarm
|
None
|
|
Speed alarm
|
None
|
|
Speed shut off
|
None
|
|
Unbalanced load
|
None
|
|
Soft start feature
|
None
|
|
RS232 interface
|
None
|
|
Recorder output
|
None
|
|
Motor
|
Solid-state, DC brushless motor
|
|
Power rating
|
120V, 50/60Hz, 4.5A
|
|
Power consumption
|
500W
|
|
Platform dimensions (W x D)
|
11 x 13 in (28 x 33 cm)
|
|
Overall dimensions (W x D x H)
|
14.1 x 27.2 x 15.8 in (35.8 x 69.1 x 40.1 cm)
|