- Beckman Coulter J6-MI High-Capacity Refrigerated Floor Centrifuge (360291)
- Beckman Coulter JS-5.2 Swinging Bucket Rotor with Windshield
- x4 Beckman 5.2/4.0 Rotor Buckets
- Rotor Tie Down
- Attached Power Cable with NEMA 6-30 Plug
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Beckman Coulter J6-MI refrigerated floor centrifuge with the following results:
- Ran to 4,000 rpm: PASSED
- Refrigeration held at -10°C: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Performed inspection; no issue found: PASSED
Notice: This centrifuge and rotor have some cosmetic blemishes. Please review the available photos to see the physical condition before purchasing.
|
Manufacturer
|
Beckman Coulter
|
|
Model
|
J6-MI
|
|
Cat. No.
|
360291
|
|
Serial No.
|
CJE18L07
|
|
Mfg. date
|
2018-11-19
|
|
Set speed
|
100 to 6000 rpm
|
|
Speed control
|
±20 rpm of set speed
|
|
Maximum g-force
|
6840 x g
|
|
Maximum capacity
|
6 liters
|
|
Temperature setting range
|
-20°C to +40°C
|
|
Temperature control
|
±2°C of set temperature after equilibration
|
|
Accel/Decel profiles
|
9/9
|
|
Power rating
|
208V, 60Hz, 25A
|
|
Dimensions (W x D x H)
|
28 x 36 x 50 in
|
|
Weight
|
555 lb (252 kg)
|
|
Manufacturer
|
Beckman Coulter
|
|
Model
|
JS-5.2
|
|
Serial No.
|
777
|
|
Maximum speed
|
5200 rpm
|
|
Critical speed range*
|
600 to 800 rpm
|
|
Maximum solution density
|
1.2 g/mL
|
|
Maximum allowable imbalance of opposing loads
|
10 grams
|
|
Relative centrifugal field at maximum speed
|
At rmax (226 mm)... 6840 × g At rav (156 mm)... 4720 × g At rmin (86 mm)... 2600 × g
|
|
Number of buckets
|
4
|
|
Nominal capacity per bucket
|
1 liter bottle, or 1 quad pack blood bag
|
|
Nominal capacity of rotor
|
4 liters, 4 blood bags, 12 microplates, 148 RIA tubes
|
|
Approximate acceleration time to maximum speed (rotor fully loaded)
|
2.5 minutes
|
|
Approximate deceleration time from maximum speed (rotor fully loaded)
|
3 minutes
|
|
Rotor and bucket material
|
Anodized aluminum
|
|
Rotor entry code for microprocessor-controlled instruments
|
5.2
|
|
Weight of fully loaded rotor
|
46 lb (21 kg)
|
*The critical speed range is the range of speeds over which the rotor shifts so as to rotate about its center of mass. Passing through the critical speed range is characterized by some vibration.