- Eppendorf 5427R Refrigerated Benchtop Microcentrifuge (5409)
- Eppendorf F-45-30-11 Fixed-Angle Rotor and Rotor Lid
- Power Cable
- 90-Day Warranty
After its initial disinfection and inspection, our technicians tested the Eppendorf 5427R refrigerated benchtop microcentrifuge with the following results:
- Ran to 14,000 rpm: PASSED
- Refrigeration held at -4°C during run: PASSED
- No excessive noise or imbalance: PASSED
- Spun down properly: PASSED
- Performed inspection; no issue found: PASSED
Manufacturer
|
Eppendorf
|
Model
|
5427R
|
Model No.
|
5409
|
Serial No.
|
5409CP002360
|
Mfg. date
|
October 2013
|
Speed range
|
100-16,220 rpm (50 rpm steps)
|
Max. RCF with fixed-angle rotor
|
25,001 x g
|
Max. RCF with swing bucket rotor
|
16,049 x g
|
Max. rotor capacity
|
48 x 1.5/2.0 mL
|
Timer
|
5s to 9:59h
|
SOFT brake function
|
yes
|
Noise level
|
<55 dB(A) with rotor FA-45-48-11
|
Cooling
|
Refrigerated
|
Temperature range
|
-11°C to +40°C
|
Power rating
|
120V, 50-60Hz, 4.6A
|
Power consumption
|
550W
|
Dimensions (W x D x H)
|
12.6 x 21.3 x 10 in (31.9 x 54 x 25.4 cm)
|
Weight
|
66.1 lb (30 kg)
|