This FluoroMax-3 is untested as it requires Datamax software that is currently not installed on the computer. Without the Datamax software, the software installed will not start up and no testing can be done.
This has been inspected and is in good condition with no damage; however, we believe that it will work without issue.
- Horiba Fluoromax 3 with FluorEssence Spectrofluorometer
We recommend reviewing the photos.
Manufacturer
|
Horiba Jobin Yvon
|
Model
|
Fluoromax 3
|
Serial Number
|
3624B
|
Place of Mfg.
|
Made in USA
|
Excitation Source
|
150-W xenon, continuous output, ozone-free lamp
|
Optics
|
All-reflective, for focusing at all wavelengths and precise imaging for microsamples.
|
Dispersion
|
4.25 nm mm–1
|
Monochromators
|
Monochromators are f/3.5 Czerny-Turner design with classically-ruled gratings and all-reflective optics, using 1200- grooves/mm gratings.
Resolution
|
0.3 nm
|
Maximum scan speed
|
200 nm s–1
|
Accuracy
|
±0.5 nm
|
Step Size
|
0.0625–100 nm
|
Range
|
0–950 nm (physical)
|
Excitation Gratings
|
330-nm blaze (220–600 nm optical range)
|
Emission
|
Emission
|
|
Sample Module
|
Removable gap-bed assembly for sampling accessory replacement.
|
Detectors
|
• Calibrated photodiode for excitation reference correction from 200–980 nm.
• Emission detector is an R928P for high sensitivity in photoncounting mode (180–850 nm). High voltage = 950 V, linearity to 2 × 106 counts s–1, < 1000 dark counts s–1.
|
Sensitivity
|
Double-distilled, de-ionized, ICP-grade water-Raman scan 2500:1 S/N at 397 nm, 5-nm bandpass, 1-s integration time, background noise first standard deviation at 450 nm. 300 000 counts s–1 using these conditions.
|
Excitation shutter
|
Computer-controlled
|
Integration time
|
0.001–160 s
|
Slit width
|
0–30 nm bandpass, continuously adjustable via host computer
|
Dimensions (W x D x H)
|
32.5 x 20.5 x 11 in.
|
Weight
|
75 lbs (34 kg)
|
Ambient temperature range
|
15–30°C (59–86°F)
|
Maximum relative humidity
|
75%
|
Electrical Specifications
|
115V, 4 A, 60 Hz, 1 PH
|