After its initial disinfection and inspection, our technicians tested the Leica CM1850 UV-3-1 microtome cryostat with the following results:
- Chamber temperature maintains -34°C: PASSED
- Sample advance: PASSED
- Chamber light: PASSED
- Disinfection (UV) system: PASSED
- Window heater: PASSED
- Handwheel brake: PASSED
- Defrost: PASSED
- Reinsulated refrigeration lines.
- Recharged refrigerant.
- Installed a new compressor.
- Installed a new fan motor.
- Cleaned undercarriage.
- Cleaned refrigeration condenser.
- Cleaned chamber with microtome removed.
- Cleaned and lubricated microtome.
- Cleaned and lubricated knife assembly.
Notice: This cryostat has some minor cosmetic blemishes. Please review the available photos to see the physical condition before purchasing.
Manufacturer
|
Leica
|
Model
|
CM1850 UV-3-1
|
Catalog No.
|
047140225
|
Serial No.
|
1323/05.2008
|
Mfg. date
|
05.2008
|
Cooling range
|
-35°C to 0°C
|
Section thickness range
|
1-60 µm From 1 to 10 µm in 1 µm intervals From 10 to 20 µm in 2 µm intervals From 20 to 60 µm in 5 µm intervals
|
Horizontal specimen feed
|
25 mm / 0.98 in
|
Vertical stroke
|
59 mm / 2.32 in
|
Maximum specimen size
|
55 x 55 mm / 2.17 x 2.17 in
|
Specimen orientation
|
8° (x-, y-, z-axis)
|
Disinfection time
|
30 or 180 min. cycle
|
Power rating
|
120V~, 60Hz
|
Power consumption
|
1440VA
|
Dimensions (W x D x H)
|
29 x 27.5 x 47.5 in
|
Weight
|
296.62 lb (~135 kg)
|