- Midmark Ritter M11 UltraClave Automatic Steam Sterilizer Autoclave (M11-022)
- Tray Rack
- Power Cable
- 90-Day Warranty
Notice: This autoclave does not come with trays.
After its initial disinfection and inspection, our technicians tested the Midmark Ritter M11 UltraClave automatic steam sterilizer with the following results:
- Powers on and initializes to “Select Cycle” screen: PASSED
- Ran cycle to 270°F for 30 minutes: PASSED
- Heating: PASSED
- Pressure: PASSED
- Filled water: PASSED
- Drained water: PASSED
Notice: This autoclave has some cosmetic blemishes. Please review the available photos to see the physical condition before purchasing.
|
Manufacturer
|
Midmark
|
|
Model
|
M11 UltraClave
|
|
Model No.
|
M11-022
|
|
Serial No.
|
V689032
|
|
Usable chamber volume
|
6.5 gal (24.6 liters)
|
|
Water reservoir capacity
|
1.4 gal (5.3 liters)
|
|
Safety standards
|
UL listed to US and Canada standards UL 61010-1, 2nd edition IEC 61010-2-040, 1st edition CAN/CSA C22.2 No. 61010-1, 2nd edition CSA C22.2 No. 61010-2-040-07, Part 2-040, 1st edition
Meets the requirements of ASME Boiler and Pressure Vessel Code. A separate (dedicated) circuit is recommended for this sterilizer.
|
|
Certifications
|
IEC 60601-1-2 Edition 4.0 2014-02 CISPR II:2009, A1:2010 Class A, Group 1 IEC 61000-3-2:2014, Part 3-2 IEC 61000-3-3:2014, Part 3-3 ANSI/AAMI ST55:2010(R)2014 ANSI/AAMI/IEC 62304
|
|
Installation
|
Use on a level, water-resistant material such as laminate, stainless-steel or stone surface. Allow 2.0" (5.08 cm) on sides and rear, and 5" (12.7 cm) above to allow space to fill the reservoir and change the printer paper roll.
|
|
Trays (W x D x H)
|
Two large – 9 x 15 x 1.1 in (22.8 x 38 x 2.9 cm) Two small – 6.6 x 15 x 1.1 in (16.8 x 38 x 2.9 cm)
|
|
Chamber dimensions (W x D)
|
11 x 18 in (28 x 45.7 cm)
|
|
Exterior dimensions (W x D x H)
|
17.8 x 23.8 x 18.1 in (45.2 x 60.5 x 50 cm)
|
|
Power rating
|
115VAC, 50/60Hz, 12A
|
|
Weight
|
99 lb (44.9 kg)
|