After your instrument arrives at our facility, it will be inspected and tested to verify that it is free of malfunction. Unless stated otherwise, we are confident it will perform at manufacturer specifications. If you receive this instrument and it does not meet your required specifications, your purchase will be protected by our warranty.
In Good Cosmetic Condition or Better
The photos on display are stock photos and do not directly represent the instrument you will receive. However, we guarantee your instrument will be in as good cosmetic condition or better when it arrives at your location. For reference, used equipment in good cosmetic condition commonly has a few scratches and scuffs present.
Included Items:
- Sanyo MCO-19AIC (UV) CO2 Incubator
- Power Cable
Specifications
Manufacturer
|
Sanyo
|
Model
|
MCO-19AIC
|
External dimensions
|
620 x 710 x 900 mm (W x D x H)
|
Internal dimensions
|
490 x 523x 665 mm (W x D x H)
|
Interior volume
|
170 L
|
Access port
|
Inner diameter: 30 mm, On the back side
|
Insulation
|
Rigid polyurethane foamed-in place
|
Heating system
|
DHA system (heater jacket + air jacket system)
|
Heater
|
314 W
|
Humidifying system
|
Natural evaporation with humidifying pan
|
Temperature controller
|
PID control system
|
Temperature display
|
Digital display
|
CO2 controller
|
PID control system
|
CO2 density display
|
Digital display
|
Air circulation
|
Fan assisted
|
Temperature control range
|
+5°C above ambient to 50°C (in a 5°C to 35°C ambient).
|
Temperature control uniformity
|
±0.25°C in 25°C ambient, setting 37°C, 5% CO2, no load.
|
C02 control range and deviation
|
0% to 20%, ±0.15% in 25°C ambient, setting 37°C, 5% CO2, no load.
|
C02 sensor platform
|
Ceramic based, single beam Infrared sensor, with dual wavelength measurement for continuous auto-zero calibration
|
Air filter
|
0.3 Pm, Efficiency: 99.97% or higher
|
UV lamp
|
4W X 1 (ozone-free emission)
|
Water level sensor
|
Optical
|
Alarms
|
Automatic set temperature alarm, Automatic set CO2 density alarm,
Upper limit temperature alarm, Door alarm, Low humidifying water alarm, UV lamp failure alarm (for UV equipped model only)
|
Remote alarm contacts
|
Allowable contact capacity: 30 VDC, 2 A
|
CO2 inlet connection
|
4 to 6 mm diameter tube
|
CO2 inlet pressure
|
0.03 MPaG (0.3 kgf/cm2 G, 4.3 psiG)
|
Weight
|
93 kg
|